ANALYSIS OF THE MORPHOLOGY OF THE UNIPOLAR ENDOCARDIAL PACED EVOKED-RESPONSE

被引:12
作者
BROUWER, J
NAGELKERKE, D
DEJONGSTE, MJL
BOUTE, W
DENHEIJER, P
LIE, KI
机构
[1] STATE UNIV GRONINGEN HOSP,THORAXCTR,DEPT CARDIOL,9713 EZ GRONINGEN,NETHERLANDS
[2] VITATRON MED BV,DIEREN,NETHERLANDS
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1990年 / 13卷 / 03期
关键词
electrode characteristics; evoked response; morphology; rate responsive pacing; stimulus T interval; T wave sensing;
D O I
10.1111/j.1540-8159.1990.tb02044.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In a retrospective study we analyzed the unipolar endocardial evoked response signal (ERS) of 103 patients prior to pacemaker implantation. The objective of this study was to give a complete description of the ERS morphology and to evaluate influences on this morphology of both various electrode characteristics and pacing rate. In addition, spontaneous endocardial signals were studied. The results demonstrate that acute leads had both higher R wave and T wave amplitudes and a faster downslope of the T wave. In the acute leads those with porous titanium carbon coated tips showed a more pronounced T wave. Pacing rate influences the R wave amplitude and the stimulus to T wave interval. Both stimulus to maximum and stimulus to minimum T wave interval show an exponential correlation with the stimulus interval. The interval between maximum and minimum of the T wave and the absolute amplitude of the T wave are not influenced by rate. Although there were significant correlations of the spontaneous endocardial signal with the ERS, the predictive value of the spontaneous signal for the ERS morpology is low. Prospective studies will be necessary to confirm the findings in this study. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:302 / 313
页数:12
相关论文
共 27 条
  • [1] AUTODIAGNOSTIC PACEMAKER
    AUERBACH, AA
    FURMAN, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1979, 2 (01): : 58 - 68
  • [2] NONLINEAR RELATIONSHIP BETWEEN PACING AND EVOKED QT INTERVALS
    BAIG, MW
    BOUTE, W
    BEGEMANN, M
    PERRINS, EJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (06): : 753 - 759
  • [3] Bazett HC, 1920, HEART-J STUD CIRC, V7, P353
  • [4] RELIABILITY OF EVOKED ENDOCARDIAL T-WAVE SENSING IN 1,500 PACEMAKER PATIENTS
    BOUTE, W
    DERRIEN, Y
    WITTKAMPF, FHM
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (06): : 948 - 953
  • [5] BOUTE W, 1987, VITATEXT JUN
  • [6] BOUTE W, 1986, PACE S, V4, P47
  • [7] DEJONGSTE MJL, 1985, CARDIAC PACING, P841
  • [8] DENHEIJER P, 1989, PACE, V12, P805
  • [9] THE VENTRICULAR ENDOCARDIAL PACED EVOKED-RESPONSE
    DONALDSON, RM
    RICKARDS, AF
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (02): : 253 - 259
  • [10] RATE RESPONSIVE PACING USING THE EVOKED QT PRINCIPLE - A PHYSIOLOGICAL ALTERNATIVE TO ATRIAL SYNCHRONOUS PACEMAKERS
    DONALDSON, RM
    RICKARDS, AF
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1983, 6 (06): : 1344 - 1349