PROBLEMS RELATED TO THE ENDOTRACHEAL-TUBE - AN ANALYSIS OF 2000 INCIDENT REPORTS

被引:92
作者
SZEKELY, SM
WEBB, RK
WILLIAMSON, JA
RUSSELL, WJ
机构
[1] UNIV ADELAIDE,DEPT ANAESTHESIA & INTENS CARE,HYPERBAR MED UNIT,ADELAIDE,SA 5001,AUSTRALIA
[2] ROYAL ADELAIDE HOSP,ADELAIDE,SA 5000,AUSTRALIA
关键词
INTUBATION; ENDOBRONCHIAL; ESOPHAGEAL; ENDOTRACHEAL TUBE; OBSTRUCTION; CATHETER MOUNT; CUFF PROBLEMS; CUFF LEAK;
D O I
10.1177/0310057X9302100520
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The first 2000 incidents reported to the Australian Incident Monitoring Study were analysed with respect to problems with the endotracheal tube; 189 (9%) were reported. The most common problem was endobronchial intubation which accounted for 42% of these 189 reports, endobronchial intubation was the most common cause of arterial desaturation in the 2000 incidents. Obstructions and oesophageal intubation each accounted for 18% of the 189 problems with tubes. The remainder was made up of disconnections and leaks (7% each), misplacements other than endobronchial or oesophageal (4%), inappropriate choice of tube (3%), cuff herniation (1%), failure to deflate the cuff and foreign body in the tube (0.5% each). The pulse oximeter and capnograph first detected 58% of these incidents; a further 25% were detected clinically. The pulse oximeter is the ''front-line'' monitor for endobronchial intubation, and the capnograph the ''front-line'' monitor for oesophageal intubation, disconnection and obstruction. Recommendations are made for how to prevent problems and how to determine the nature of those that do occur.
引用
收藏
页码:611 / 616
页数:6
相关论文
共 14 条
[1]   THE OXYGEN ANALYZER - APPLICATIONS AND LIMITATIONS - AN ANALYSIS OF 2000 INCIDENT REPORTS [J].
BARKER, L ;
WEBB, RK ;
RUNCIMAN, WB ;
VANDERWALT, JH .
ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) :570-574
[2]   ADVERSE RESPIRATORY EVENTS IN ANESTHESIA - A CLOSED CLAIMS ANALYSIS [J].
CAPLAN, RA ;
POSNER, KL ;
WARD, RJ ;
CHENEY, FW .
ANESTHESIOLOGY, 1990, 72 (05) :828-833
[3]   ESOPHAGEAL INTUBATION - AN ANALYSIS OF 2000 INCIDENT REPORTS [J].
HOLLAND, R ;
WEBB, RK ;
RUNCIMAN, WB .
ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) :608-610
[4]   CARDIAC-ARREST - AN ANALYSIS OF 2000 INCIDENT REPORTS [J].
MORGAN, CA ;
WEBB, RK ;
COCKINGS, J ;
WILLIAMSON, JA .
ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) :626-637
[5]   ACCIDENTAL INTUBATION OF THE ESOPHAGUS [J].
POLLARD, BJ ;
JUNIUS, F .
ANAESTHESIA AND INTENSIVE CARE, 1980, 8 (02) :183-186
[6]   CRISIS MANAGEMENT - VALIDATION OF AN ALGORITHM BY ANALYSIS OF 2000 INCIDENT REPORTS [J].
RUNCIMAN, WB ;
WEBB, RK ;
KLEPPER, ID ;
LEE, R ;
WILLIAMSON, JA ;
BARKER, L .
ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) :579-592
[7]   THE PULSE OXIMETER - APPLICATIONS AND LIMITATIONS - AN ANALYSIS OF 2000 INCIDENT REPORTS [J].
RUNCIMAN, WB ;
WEBB, RK ;
BARKER, L ;
CURRIE, M .
ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) :543-550
[8]   PROBLEMS WITH VENTILATION - AN ANALYSIS OF 2000 INCIDENT REPORTS [J].
RUSSELL, WJ ;
WEBB, RK ;
VANDERWALT, JH ;
RUNCIMAN, WB .
ANAESTHESIA AND INTENSIVE CARE, 1993, 21 (05) :617-620
[9]   ENDOTRACHEAL-TUBE MISPLACEMENT [J].
STIRT, JA .
ANAESTHESIA AND INTENSIVE CARE, 1982, 10 (03) :274-276
[10]   HUMAN MISADVENTURE IN ANESTHESIA [J].
UTTING, JE ;
GRAY, TC ;
SHELLEY, FC .
CANADIAN ANAESTHETISTS SOCIETY JOURNAL, 1979, 26 (06) :472-478