OPTIMIZATION AND VALIDATION OF ANALYTICAL CONDITIONS FOR BOVINE SERUM-ALBUMIN USING CAPILLARY ELECTROPHORESIS

被引:23
|
作者
PANDE, PG [1 ]
NELLORE, RV [1 ]
BHAGAT, HR [1 ]
机构
[1] UNIV MARYLAND,SCH PHARM,DEPT PHARMACEUT,BALTIMORE,MD 21201
关键词
D O I
10.1016/0003-2697(92)90146-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A quick and reproducible capillary electrophoresis method was optimized and validated for the assay of bovine serum albumin (BSA). The effects of various parameters such as pH of buffer, concentration of buffer, capillary dimensions, use of coated capillaries, and additives such as surfactants and protein solubilizers were evaluated. The capillary coatings or additives did not give any advantage in reducing the surface adsorption of BSA on the capillary walls. The optimized conditions include use of borate buffer, pH 8.5 having a concentration of 150 mm in a 27 cm capillary with an aperture window of 100 × 200 μm for detection. The optimized method for the detection of BSA was validated. The interday and intraday coefficient of variation was not greater than 7.59% at BSA concentrations of 25-1000 μg/ml. The method developed was reproductible and accurate. © 1992.