LOW-LEVEL ELECTROCHEMICAL DETECTION OF GLUCOSE-OXIDASE AND A GLUCOSE-OXIDASE CONJUGATE

被引:1
作者
ALLEN, RM [1 ]
BENNETTO, HP [1 ]
LEWIS, IAS [1 ]
MASON, JR [1 ]
WESTRA, AR [1 ]
机构
[1] UNIV LONDON KINGS COLL,DEPT CHEM,LONDON W8 7AH,ENGLAND
关键词
GLUCOSE OXIDASE; PLATINIZED CARBON PAPER ELECTRODE; AMPEROMETRY; HYDROGEN PEROXIDE; GLUCOSE OXIDASE CONJUGATE; ENZYME LABEL;
D O I
10.1016/0956-5663(95)96938-U
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
An amperometric sensor for glucose oxidase is described. Partial optimization of the working electrode potential at 0.4 V vs. Ag/AgCl, 0.1 M KCl at a glucose concentration of 30 mM allows a low detection limit of 50 pM glucose oxidase within 25 min. The sensor exhibits a high degree of precision and reproducibility, and the same electrode may be re-used for many successive quantitative determinations. A mechanism for the sensor is proposed, in which hydrogen peroxide generated by glucose oxidase in bulk solution is oxidized at the working electrode, generating a current response that can be related to the glucose oxidase concentration. Conjugation of glucose oxidase with a coumarin-based pesticide analogue does not affect the amperometric response.
引用
收藏
页码:621 / 631
页数:11
相关论文
共 7 条
  • [1] Allen, Westra, Bennetto, Lewis, Mason, Development of immuno-concentration and electrochemical detection systems, Proc. 2nd European workshop on Biosensors for Environmental Monitoring, pp. 3-8, (1994)
  • [2] Aubree-Lecat, Hervagault, Delacour, Beaude, Bourdillon, Remy, Direct electrochemical determination of glucose oxidase in biological samples, Anal. Biochem., 178, pp. 427-430, (1989)
  • [3] Bennetto, DeKeyser, Delaney, Koshy, Mason, Mourla, Razack, Stirling, Thurston, Anderton, Mullen, A glucose oxidase electrode for amperometric determination of glucose, Int. Ind. Biotech., 8, pp. 5-10, (1988)
  • [4] Bennetto, DeKeyser, Delaney, Koshy, Mason, Razack, Stirling, Thurston, An amperometric biosensor for laboratory determination of glucose, Int. Anal., 1, pp. 22-27, (1987)
  • [5] Cass, Davis, Francis, Hill, Aston, Higgins, Plotkin, Scott, Turner, Ferrocene-mediated enzyme electrode for amperometric determination of glucose, Anal. Chem., 56, pp. 667-671, (1984)
  • [6] Wilson, Turner, Glucose oxidase: an ideal enzyme, Biosensors & Bioelectr., 7, pp. 165-185, (1992)
  • [7] Zilkha, Koshy, Obrenovitch, Bennetto, Symon, Amperometric biosensors for on-line monitoring of extracellular glucose and glutamate in the brain, Analytical Letters, 27, 3, pp. 453-473, (1994)