Diagnostic accuracy of the iCare rebound tonometer compared to the Perkins applanation tonometer in assessing intraocular pressure in rural patients

被引:6
作者
Li, Yifan [4 ]
Carpenter, Christopher R. [5 ]
Nicholson, Kathryn [6 ]
Milne, William Ken [1 ,2 ,3 ]
机构
[1] South Huron Hosp Assoc, 24 Huron St W, Exeter, ON N0M 1S2, Canada
[2] Univ Western Ontario, Schulich Sch Med & Dent, Dept Med, Goderich, ON, Canada
[3] South Huron Hosp Assoc, Dept Emergency Med, Exeter, ON, Canada
[4] Univ Western Ontario, Windsor Program, Schulich Sch Med & Dent, Windsor, ON, Canada
[5] Washington Univ, St Louis Sch Med, Div Emergency Med, St Louis, MO USA
[6] Univ Western Ontario, Schulich Sch Med & Dent, Ctr Studies Family Med, Dept Epidemiol & Biostat, London, ON, Canada
关键词
glaucoma; iCare rebound tonometer; IOP; Perkins applanation tonometer; rural; screening;
D O I
10.1515/dx-2015-0026
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Vision health is recognized as a critical unmet need in North America. The ocular morbidity associated with glaucoma results from increased intraocular pressure (IOP) and early detection is crucial for the management of glaucoma. Our objective was to find a diagnostically accurate screening tool for intraocular hypertension that can be used in rural communities. We sought to validate the diagnostic accuracy of the iCare rebound tonometer against the gold standard Perkins applanation tonometer (PAT) in measuring IOP. Methods: Patients from two rural communities in Ontario, Canada visiting their optometrists for routine appointments had their IOP measured by a non-contact tonometer (NCT), an iCare rebound tonometer, and a Perkins applanation tonometer (PAT). Values of sensitivity, specificity, and likelihood ratios for a positive and negative result were calculated for the iCare and the NCT. Results: Complete data was collected from 209 patients. Overall, the iCare tonometer had high levels of validity, as compared to the gold standard PAT. The iCare tonometer displayed excellent sensitivity of 98.3% (90-99%, 95% CI) and excellent negative likelihood ratio of 0.024 (0.0088-0.066, 95% CI) which is useful for ruling out intraocular hypertension. Conclusions: The iCare tonometer is a reasonably valid tool for detecting elevated IOP. Its ease of use, simplicity, and accessibility makes it a good screening tool to improve eye health in rural areas.
引用
收藏
页码:227 / 234
页数:8
相关论文
共 50 条
  • [21] Effectiveness of the ICare rebound tonometer in patients with overestimated intraocular pressure due to tight orbit syndrome
    Lee, Y. K.
    Lee, J. Y.
    Moon, J. I.
    Park, M. H.
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2014, 58 (06) : 496 - 502
  • [22] Comparison of the Icare ic100 Rebound Tonometer and the Goldmann Applanation Tonometer in 1,000 Eyes
    Subramaniam, Arjun Gopal
    Allen, Penelope
    Toh, Tze'Yo
    OPHTHALMIC RESEARCH, 2021, 64 (02) : 321 - 326
  • [23] Reliability and reproducibility of introcular pressure (IOP) measurement with the Icare® Home rebound tonometer (model TA022) and comparison with Goldmann applanation tonometer in glaucoma patients
    Valero, B.
    Fenolland, J. -R.
    Rosenberg, R.
    Sendon, D.
    Mesnard, C.
    Sigaux, M.
    Giraud, J. -M.
    Renard, J. -P.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2017, 40 (10): : 865 - 875
  • [24] Evaluation of monkey intraocular pressure by rebound tonometer
    Yu, Wenhan
    Cao, Guiqun
    Qiu, Jinghua
    Liu, Xuyang
    Ma, Jia
    Li, Ni
    Yu, Man
    Yan, Naihong
    Chen, Lei
    Pang, Iok-Hou
    MOLECULAR VISION, 2009, 15 (235-36): : 2196 - 2201
  • [25] Study of Accuracy of Intraocular Pressure measured by non-contact (air puff) Tonometer confirmed by Goldmann Applanation Tonometer
    Dibaji, Mohammad
    Shaikh, Rehan Moinuddin
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2016, 10 (03): : 972 - 974
  • [26] Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects
    Kato, Yoshitake
    Nakakura, Shunsuke
    Matsuo, Naoko
    Yoshitomi, Kayo
    Handa, Marina
    Tabuchi, Hitoshi
    Kiuchi, Yoshiaki
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (02) : 687 - 696
  • [27] Agreement among Goldmann applanation tonometer, iCare, and Icare PRO rebound tonometers; non-contact tonometer; and Tonopen XL in healthy elderly subjects
    Yoshitake Kato
    Shunsuke Nakakura
    Naoko Matsuo
    Kayo Yoshitomi
    Marina Handa
    Hitoshi Tabuchi
    Yoshiaki Kiuchi
    International Ophthalmology, 2018, 38 : 687 - 696
  • [28] Comparison of ICare, dynamic contour tonometer, and ocular response analyzer with Goldmann applanation tonometer in patients with glaucoma
    Vandewalle, Evelien
    Vandenbroeck, Sofie
    Stalmans, Ingeborg
    Zeyen, Thierry
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2009, 19 (05) : 783 - 789
  • [29] Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer
    Farhood, Qasim K.
    CLINICAL OPHTHALMOLOGY, 2013, 7 : 23 - 27
  • [30] Clinical Evaluation of the New Rebound Tonometers Icare PRO and Icare ONE Compared With the Goldmann Tonometer
    Moreno-Montanes, Javier
    Martinez-de-la-Casa, Jose M.
    Sabater, Alfonso L.
    Morales-Fernandez, Laura
    Saenz, Cristina
    Garcia-Feijoo, Julian
    JOURNAL OF GLAUCOMA, 2015, 24 (07) : 527 - 532