Diagnostic accuracy of the iCare rebound tonometer compared to the Perkins applanation tonometer in assessing intraocular pressure in rural patients

被引:6
|
作者
Li, Yifan [4 ]
Carpenter, Christopher R. [5 ]
Nicholson, Kathryn [6 ]
Milne, William Ken [1 ,2 ,3 ]
机构
[1] South Huron Hosp Assoc, 24 Huron St W, Exeter, ON N0M 1S2, Canada
[2] Univ Western Ontario, Schulich Sch Med & Dent, Dept Med, Goderich, ON, Canada
[3] South Huron Hosp Assoc, Dept Emergency Med, Exeter, ON, Canada
[4] Univ Western Ontario, Windsor Program, Schulich Sch Med & Dent, Windsor, ON, Canada
[5] Washington Univ, St Louis Sch Med, Div Emergency Med, St Louis, MO USA
[6] Univ Western Ontario, Schulich Sch Med & Dent, Ctr Studies Family Med, Dept Epidemiol & Biostat, London, ON, Canada
关键词
glaucoma; iCare rebound tonometer; IOP; Perkins applanation tonometer; rural; screening;
D O I
10.1515/dx-2015-0026
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Vision health is recognized as a critical unmet need in North America. The ocular morbidity associated with glaucoma results from increased intraocular pressure (IOP) and early detection is crucial for the management of glaucoma. Our objective was to find a diagnostically accurate screening tool for intraocular hypertension that can be used in rural communities. We sought to validate the diagnostic accuracy of the iCare rebound tonometer against the gold standard Perkins applanation tonometer (PAT) in measuring IOP. Methods: Patients from two rural communities in Ontario, Canada visiting their optometrists for routine appointments had their IOP measured by a non-contact tonometer (NCT), an iCare rebound tonometer, and a Perkins applanation tonometer (PAT). Values of sensitivity, specificity, and likelihood ratios for a positive and negative result were calculated for the iCare and the NCT. Results: Complete data was collected from 209 patients. Overall, the iCare tonometer had high levels of validity, as compared to the gold standard PAT. The iCare tonometer displayed excellent sensitivity of 98.3% (90-99%, 95% CI) and excellent negative likelihood ratio of 0.024 (0.0088-0.066, 95% CI) which is useful for ruling out intraocular hypertension. Conclusions: The iCare tonometer is a reasonably valid tool for detecting elevated IOP. Its ease of use, simplicity, and accessibility makes it a good screening tool to improve eye health in rural areas.
引用
收藏
页码:227 / 234
页数:8
相关论文
共 50 条
  • [1] Comparison of Icare Rebound Tonometer and Perkins Applanation Tonometer in Community Eye Screening
    Ting, Siew Leng
    Lim, Lik Thai
    Ooi, Chor Yau
    Rahman, Md Mizanur
    ASIA-PACIFIC JOURNAL OF OPHTHALMOLOGY, 2019, 8 (03): : 229 - 232
  • [2] Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer
    Gao, Feng
    Liu, Hu
    Zhao, Qing
    Pan, Yingzhe
    EXPERIMENTAL AND THERAPEUTIC MEDICINE, 2017, 13 (05) : 1912 - 1916
  • [3] A comparison of postural and diurnal variations in intraocular pressure using the iCare rebound tonometer and Perkins applanation tonometer in admitted adults in Kenya
    Chikasirimobi, Timothy G.
    Ndinyo, Mercy
    Ondieki, Maxine N.
    Miti, Isaac
    Ejiochi-Iyoke, Ndidi J.
    Osuagwu, Uchechukwu L.
    Zeried, Ferial M.
    Agho, Kingsley
    Ogbuehi, Kelechi C.
    Mashige, Khathutshelo P.
    AFRICAN VISION AND EYE HEALTH JOURNAL, 2021, 80 (01):
  • [4] Comparison of the iCare rebound tonometer and the Perkins applanation tonometer in children under general anesthesia
    Chagny, M.
    Stolowy, N.
    Denis, D.
    Sauvan, L.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2023, 46 (09): : 1005 - 1012
  • [5] Comparability of Icare Pro Rebound Tonometer with Goldmann Applanation and Noncontact Tonometer in a Wide Range of Intraocular Pressure and Central Corneal Thickness
    Tamcelik, Nevbahar
    Atalay, Eray
    Cicik, Erdogan
    Ozkok, Ahmet
    OPHTHALMIC RESEARCH, 2015, 54 (01) : 18 - 25
  • [6] Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma
    Shunsuke Nakakura
    Ryo Asaoka
    Etsuko Terao
    Yuki Nagata
    Yasuko Fukuma
    Satomi Oogi
    Miku Shiraishi
    Yoshiaki Kiuchi
    Eye and Vision, 8
  • [7] Evaluation of rebound tonometer iCare IC200 as compared with IcarePRO and Goldmann applanation tonometer in patients with glaucoma
    Nakakura, Shunsuke
    Asaoka, Ryo
    Terao, Etsuko
    Nagata, Yuki
    Fukuma, Yasuko
    Oogi, Satomi
    Shiraishi, Miku
    Kiuchi, Yoshiaki
    EYE AND VISION, 2021, 8 (01)
  • [8] Comparison of Intraocular Pressure Measurements and Assessment of Intraobserver and Interobserver Reproducibility With the Portable ICare Rebound Tonometer and Goldmann Applanation Tonometer in Glaucoma Patients
    Salim, Sarwat
    Du, Haiming
    Wan, Jim
    JOURNAL OF GLAUCOMA, 2013, 22 (04) : 325 - 329
  • [9] Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range
    Munkwitz, S.
    Elkarmouty, A.
    Hoffmann, E. M.
    Pfeiffer, N.
    Thieme, H.
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2008, 246 (06) : 875 - 879
  • [10] Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients
    Brusini, P
    Salvetat, ML
    Zeppieri, M
    Tosoni, C
    Parisi, L
    JOURNAL OF GLAUCOMA, 2006, 15 (03) : 213 - 217