BACKDIFFUSION OR BICARBONATE MAY STIMULATE COMPLEMENT ACTIVATION DURING HEMODIALYSIS WITH LOW-FLUX MEMBRANES

被引:7
|
作者
LUNDBERG, L [1 ]
STEGMAYR, BG [1 ]
WEHLE, B [1 ]
机构
[1] HUDDINGE UNIV HOSP,DEPT NEPHROL,STOCKHOLM,SWEDEN
来源
关键词
HEMODIALYSIS; BACKDIFFUSION; COMPLEMENT ACTIVATION; C3D; DIALYSATE; CELLULOSE ACETATE; HEMOPHAN(R);
D O I
10.1177/039139889401700301
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Backdiffusion of dialysate during haemodialysis with low-flux membranes and the use of bicarbonate dialysatebase, may increase the risk for contamination. The influence on the complement system was studied by altering the flux of acetate or bicarbonate dialysate base across the membrane. Eight patients were dialysed with a transmembrane pressure of 100 mm Hg (group I) during the first 60 min to standardize the ultrafiltration (UF) and acetate as dialysate. In eight other patients (group II) the UF was ''set at zero'' ml during the first 60 min using an FCM 10-1 monitor (Gambro) and bicarbonate as base. The groups were dialysed three times on two hollow-fiber membranes made of Hemophan(R) and cellulose acetate (CA). Blood samples were taken at 0, 15, 60 and 180 min, and analysed for plasma protein, haematocrit and complement C3d. In group II there was a reduction in plasma protein concentration at 15 and 60 min (p<0.002) for Hemophan and at 60 min (p<0.01) using CA. C3d was increased at 15 min for both filters (p<0.03). The reduction of protein in group II was followed by changes in the haematocrit, indicating a backdiffusion of dialysate, which may contribute to the concomittant increase in C3d.
引用
收藏
页码:131 / 136
页数:6
相关论文
共 50 条
  • [1] High-Flux and Low-Flux Membranes in Hemodialysis
    Oshvandi, Khodayar
    Kavyannejad, Rasol
    Borzuo, Sayed Reza
    Gholyaf, Mahmoud
    NURSING AND MIDWIFERY STUDIES, 2014, 3 (03)
  • [2] Pyrogen transfer across high- and low-flux hemodialysis membranes
    Weber, V
    Linsberger, I
    Rossmanith, E
    Weber, C
    Falkenhagen, D
    ARTIFICIAL ORGANS, 2004, 28 (02) : 210 - 217
  • [4] Effects of Low-Flux and High-Flux Dialysis Membranes on Erythropoietin Responsiveness in Hemodialysis Patients
    Abdelhamid, Walid A. R.
    Soliman, Mohamed M.
    El-Hameed, Ayman R. A.
    INTERNATIONAL JOURNAL OF NEPHROLOGY, 2022, 2022
  • [5] Phosphorus Removal in Low-Flux Hemodialysis, High-Flux Hemodialysis, and Hemodiafiltration
    Svara, Frantisek
    Lopot, Frantisek
    Valkovsky, Ivo
    Pecha, Ondrej
    ASAIO JOURNAL, 2016, 62 (02) : 176 - 181
  • [6] COMPLEMENT ACTIVATION BY HEMODIALYSIS MEMBRANES
    KIM, Y
    MICHAEL, AF
    KJELLSTRAND, CM
    MAUER, SM
    DIALYSIS & TRANSPLANTATION, 1982, 11 (04) : 265 - &
  • [7] Clearance of iopromide during haemodialysis with high- and low-flux membranes
    Matzkies, FK
    Tombach, B
    Kisters, K
    Schuhmann, G
    Hohage, H
    Schaefer, RM
    ACTA RADIOLOGICA, 1999, 40 (02) : 220 - 223
  • [8] COMPLEMENT ACTIVATION DURING HEMODIALYSIS - COMPARISON OF CUPROPHANE AND POLYACRYLONITRILE MEMBRANES
    CHENOWETH, DE
    CHEUNG, AK
    HENDERSON, LW
    CLINICAL RESEARCH, 1982, 30 (02): : A444 - A444
  • [9] Effects of low-flux and high-flux hemodialysis on the survival of elderly maintenance hemodialysis patients
    Huang, Wanqing
    Bai, Jiuxu
    Zhang, Yanping
    Qiu, Dongxia
    Wei, Lin
    Zhao, Chen
    Ren, Zhuo
    Wang, Qian
    Ren, Kaiming
    Cao, Ning
    RENAL FAILURE, 2024, 46 (01)
  • [10] EFFECTS OF COMPLEMENT ACTIVATION BY HEMODIALYSIS MEMBRANES