CLINICAL-EVALUATION OF A NEW SINGLE-PASS LEAD VDD PACING SYSTEM

被引:33
作者
OVSYSHCHER, IE [1 ]
KATZ, A [1 ]
BONDY, C [1 ]
机构
[1] BEN GURION UNIV NEGEV,FAC HLTH SCI,BEER SHEVA,ISRAEL
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 11期
关键词
SINGLE PASS LEAD; VDD PACING; ATRIAL SENSING;
D O I
10.1111/j.1540-8159.1994.tb03763.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Twenty-five patients with second- to third-degree AV block and normal sinus function (16 males, mean age 60 +/- 18; range 15-78 years) underwent implantation of VVD pacemakers (THERA VDD, Medtronic, Inc.) with a single pass (SP) lead. Results: During implantation the mean amplitude of the atrial (A) signal was 3.9 +/- 1.4 mV (range 2.0-7.8 mV). Stable, acceptable A-signals during implantation were usually observed in the mid- or lower part of the right atrium. The lead tip electrical parameters were not compromised in any patient in order to obtain an acceptable A-signal. To verify VDD device function, patients underwent pacing system analysis on the second day and again 1, 3, and 6 months after implantation. Acute and chronic electrical measurements in the ventricle were similar to those with regular steroid leads. During follow-up tests, stable atrial sensing (A greater than or equal to 0.7 mV) was found in all but one patient (in whom A was 0.25-0.5 mV and an intermittent loss of atrial sensing occurred). There was no difference between serial measurements of A-signal amplitudes on the second day or 1, 3, and 6 months after implantation: 1.9 +/- 1.3 mV, 1.5 +/- 0.6 mV, 1.3 +/- 0.8 mV, and 1.5 +/- 1.1 mV, respectively. The mean implantation time was 54.0 +/- 17 minutes and the mean fluoroscopy time was 3.2 +/- 1.3 minutes. Conclusions: SP lead VDD pacing is reliable and easy to manage with dependable atrial sensing and ventricular pacing. The significant reduction in atrial postimplantation amplitude is related to the different techniques used for measuring acute and chronic atrial signals.
引用
收藏
页码:1859 / 1864
页数:6
相关论文
共 17 条
[1]   ITALIAN MULTICENTER STUDY ON A SINGLE LEAD VDD PACING SYSTEM USING A NARROW ATRIAL DIPOLE SPACING [J].
ANTONIOLI, GE ;
ANSANI, L ;
BARBIERI, D ;
GUARDIGLI, G ;
PERCOCO, GF ;
TOSELLI, T .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1890-1893
[2]  
ANTONIOLI GE, 1993, NEW PERSPECTIVES CAR, V3, P359
[3]   TOWARD OPTIMIZING THE DETECTION OF ATRIAL DEPOLARIZATION WITH FLOATING BIPOLAR ELECTRODES [J].
BROWNLEE, RR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (03) :431-442
[4]   CLINICAL-EVALUATION OF VDD PACING WITH A UNIPOLAR SINGLE-PASS LEAD [J].
CORNACCHIA, D ;
FABBRI, M ;
MARESTA, A ;
GRASSI, G ;
VAIANI, P .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04) :604-618
[5]   EUROPEAN MULTICENTER PROSPECTIVE FOLLOW-UP-STUDY OF 1,002 IMPLANTS OF A SINGLE LEAD VDD PACING SYSTEM [J].
CRICK, JCP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :1742-1744
[6]   A MULTICENTER EVALUATION OF A SINGLE-PASS LEAD VDD PACING SYSTEM [J].
CURZIO, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (03) :434-442
[7]   AMPLITUDE AND DIRECTION OF ATRIAL DEPOLARIZATION USING A MULTIPOLAR FLOATING CATHETER - PRINCIPLES FOR A SINGLE LEAD VDD PACING [J].
FLAMMANG, D ;
RENIRIE, L ;
BEGEMANN, M ;
VANKRIEKEN, F .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (06) :1040-1048
[8]  
FURMAN S, 1991, PACE, V14, P693
[9]  
FURMAN S, 1989, PACE, V12, P677
[10]   INITIAL CLINICAL-EXPERIENCE WITH A SINGLE PASS VDDR PACING SYSTEM [J].
LAU, CP ;
TAI, YT ;
LI, JP ;
CHUNG, FL ;
SUNG, S ;
YAMAMOTO, A .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1894-1900