VALIDATION OF AN MRI/PET LANDMARK REGISTRATION METHOD USING 3D SIMULATED PET IMAGES AND POINT SIMULATIONS

被引:37
作者
NEELIN, P
CROSSMAN, J
HAWKES, DJ
MA, Y
EVANS, AC
机构
[1] MONTREAL NEUROL INST,MCCONNELL BRAIN IMAGING CTR,MONTREAL W5A 2B4,PQ,CANADA
[2] UMDS,GUYS HOSP,DIV RADIOL SCI,IMAGE PROC GRP RADIOL SCI,LONDON,ENGLAND
关键词
POSITRON EMISSION TOMOGRAPHY (PET); MAGNETIC RESONANCE IMAGING; IMAGE REGISTRATION; IMAGE CORRELATION; MULTIMODALITY IMAGING; PET SIMULATION;
D O I
10.1016/0895-6111(93)90028-L
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Three-dimensional (3D) simulated PET images generated from MRI were used to validate a multimodality registration technique based on the identification of internal anatomical landmarks. In addition, point-based simulations were compared with registration datasets acquired over 3 yr of routine use of the technique. Registration errors were found to range from 1.0 mm at the brain centre to 2.8 mm in each dimension at the brain surface.
引用
收藏
页码:351 / 356
页数:6
相关论文
共 12 条
[1]  
ALPERT NM, 1990, J NUCL MED, V31, P1717
[2]   MRI-PET CORRELATION IN 3 DIMENSIONS USING A VOLUME-OF-INTEREST (VOI) ATLAS [J].
EVANS, AC ;
MARRETT, S ;
TORRESCORZO, J ;
KU, S ;
COLLINS, L .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1991, 11 (02) :A69-A78
[3]   PERFORMANCE EVALUATION OF THE PC-2048 - A NEW 15-SLICE ENCODED-CRYSTAL PET SCANNER FOR NEUROLOGICAL STUDIES [J].
EVANS, AC ;
THOMPSON, CJ ;
MARRETT, S ;
MEYER, E ;
MAZZA, M .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (01) :90-98
[4]  
EVANS AC, 1989, P SOC PHOTO-OPT INS, V1092, P264
[5]  
Golub G E, 1986, MATRIX COMPUTATIONS
[6]  
KAMBER M, 1991, THESIS CONCORDIA U
[7]   A NEW HEADHOLDER FOR PET, CT, AND NMR IMAGING [J].
KEARFOTT, KJ ;
ROTTENBERG, DA ;
KNOWLES, RJR .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1984, 8 (06) :1217-1220
[8]  
MA Y, 1992, P IEEE EMBS 3D ADV I, P141
[9]   ACCURATE 3-DIMENSIONAL REGISTRATION OF CT, PET, AND OR MR IMAGES OF THE BRAIN [J].
PELIZZARI, CA ;
CHEN, GTY ;
SPELBRING, DR ;
WEICHSELBAUM, RR ;
CHEN, CT .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1989, 13 (01) :20-26
[10]  
PELIZZARI CA, 1991, P ANN INT C IEEE ENG, V13, P221