GUIDEWIRE TRANSECTION DURING ROTATIONAL CORONARY ATHERECTOMY DUE TO GUIDE CATHETER DISLODGMENT AND WIRE KINKING

被引:15
作者
FOSTERSMITH, K [1 ]
GARRATT, KN [1 ]
HOLMES, DR [1 ]
机构
[1] MAYO CLIN & MAYO FDN,DIV CARDIOVASC DIS & INTERNAL MED,ROCHESTER,MN 55905
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1995年 / 35卷 / 03期
关键词
GUIDEWIRE FRACTURE; ROTATIONAL CORONARY ATHERECTOMY; FOREIGN BODY RETRIEVAL;
D O I
10.1002/ccd.1810350313
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Rotational coronary atherectomy is an effective treatment for calcified ostial lesions. We report a case of guidewire transection during rotational atherectomy of a right coronary artery ostial stenosis. Guide catheter dislodgement appeared to have caused prolapse and kinking of the guidewire. Advancement of the burr over the kinked wire resulted in transection. The wire fragment was retrieved successfully using an inflated fixed-wire balloon catheter. This report illustrates the importance of excellent coaxial guide catheter alignment with rotational atherectomy and suggests that operators be vigilant to possible damage to the radiolucent rotational atherectomy guidewire. (C) 1995 Wiley-Liss, Inc.
引用
收藏
页码:224 / 227
页数:4
相关论文
共 24 条
[1]   COMPLICATIONS ASSOCIATED WITH THE GUIDE WIRE IN PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
ARCEGONZALEZ, JM ;
SCHWARTZ, L ;
GANASSIN, L ;
HENDERSON, M ;
ALDRIDGE, H .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (01) :218-221
[2]   SURGICAL REMOVAL OF THE INTRACORONARY PORTION OF A FRACTURED ANGIOPLASTY GUIDEWIRE [J].
BACHENHEIMER, LC ;
GREEN, CE ;
ROSING, DR ;
WALLACE, RB .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (11) :946-947
[3]   FRACTURED AND RETAINED GUIDE-WIRE FRAGMENT DURING CORONARY ANGIOPLASTY - UNFORESEEN LATE SEQUELAE [J].
DOOREY, AJ ;
STILLABOWER, M .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 20 (04) :238-240
[4]   RETRIEVAL OF A FRACTURED USCI PROBE TIP FROM A DISEASED CORONARY-ARTERY USING ANOTHER FIXED-WIRE BALLOON CATHETER, THE CORDIS ORION [J].
FELDMAN, RL ;
TRICE, WA ;
HENNEMANN, WW ;
FURST, A .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 19 (04) :257-263
[5]  
GHOSH PK, 1989, J THORAC CARDIOV SUR, V97, P467
[6]   REMOVAL OF RETAINED INTRACORONARY PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY EQUIPMENT BY A PERCUTANEOUS TWIN GUIDEWIRE METHOD [J].
GURLEY, JC ;
BOOTH, DC ;
HIXSON, C ;
SMITH, MD .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 19 (04) :251-256
[7]   RETAINED PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY EQUIPMENT COMPONENTS AND THEIR MANAGEMENT [J].
HARTZLER, GO ;
RUTHERFORD, BD ;
MCCONAHAY, DR .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 60 (16) :1260-1264
[8]   GUIDEWIRE SNAP CAUSING LEFT MAIN CORONARY-OCCLUSION DURING CORONARY ANGIOPLASTY [J].
KELTAI, M ;
BARTEK, I ;
BIRO, V .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1986, 12 (05) :324-326
[9]   FRACTURE AND DISLODGMENT OF FLOPPY GUIDEWIRE DURING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
KHONSARI, S ;
LIVERMORE, J ;
MAHRER, P ;
MAGNUSSON, P .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 58 (09) :855-856
[10]  
KRONE RJ, 1986, CATHET CARDIOVASC DI, V2, P409