ELIMINATION OF INTERFERENCES IN FLOW-INJECTION AMPEROMETRIC DETERMINATION OF GLUCOSE IN BLOOD-SERUM USING IMMOBILIZED GLUCOSE-OXIDASE

被引:19
作者
MATUSZEWSKI, W
TROJANOWICZ, M
LEWENSTAM, A
机构
[1] UNIV WARSAW,DEPT CHEM,PASTEURA 1,PL-02093 WARSAW,POLAND
[2] ABO AKAD UNIV,DEPT ANALYT CHEM,SF-20500 TURKU,FINLAND
关键词
D O I
10.1002/elan.1140020807
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow‐injection system with anodic amperometric detection for glucose determination in human blood serum samples was optimized by a comparison of various arrangements of enzyme immobilization and the elimination of interferences. The best results were obtained using a compact membrane biosensor in a wall‐jet, flow‐through detector with a platinum disk electrode covered with an evaporated Nafion layer, a polyester membrane with immobilized glucose oxidase, and another protective polyester membrane. In the optimized conditions, a linear response was observed up to 25 mM glucose with a detection limit of 200 μM glucose and a sampling rate of 120 samples/h for a 20 μL sample volume. Results of glucose determination in human serurn were compared with a Beckman Glucose Analyzer 2 and a Kone Dynamic analyzer. Copyright © 1990 VCH Publishers, Inc.
引用
收藏
页码:607 / 615
页数:9
相关论文
共 26 条
[1]  
BLAEDEL WJ, 1980, ANAL CHEM, V52, P14691
[2]   FLOW-INJECTION ANALYSIS FOR GLUCOSE AND UREA WITH ENZYME REACTORS AND ONLINE DIALYSIS [J].
GORTON, L ;
OGREN, L .
ANALYTICA CHIMICA ACTA, 1981, 130 (01) :45-53
[3]   PLATINUM-DISPERSED NAFION FILM MODIFIED GLASSY-CARBON AS AN ELECTROCATALYTIC SURFACE FOR AN AMPEROMETRIC GLUCOSE ENZYME ELECTRODE [J].
GUNASINGHAM, H ;
TAN, CB .
ANALYST, 1989, 114 (06) :695-698
[4]   GLUCOSE ENZYME ELECTRODE PREPARATION BY THE CODEPOSITION OF PLATINUM AND GLUCOSE-OXIDASE ONTO GLASSY-CARBON [J].
GUNASINGHAM, H ;
TAN, CB .
ELECTROANALYSIS, 1989, 1 (03) :223-227
[5]   CHARACTERIZATION OF PERFLUOROSULFONIC ACID POLYMER COATED ENZYME ELECTRODES AND A MINIATURIZED INTEGRATED POTENTIOSTAT FOR GLUCOSE ANALYSIS IN WHOLE-BLOOD [J].
HARRISON, DJ ;
TURNER, RFB ;
BALTES, HP .
ANALYTICAL CHEMISTRY, 1988, 60 (19) :2002-2007
[6]  
HO MH, 1984, ISA, P93
[7]   A GLUCOSE SENSOR BASED ON GLUCOSE-DEHYDROGENASE ADSORBED ON A MODIFIED CARBON ELECTRODE [J].
MARKOVARGA, G ;
APPELQVIST, R ;
GORTON, L .
ANALYTICA CHIMICA ACTA, 1986, 179 :371-379
[8]  
MASOOM M, 1984, ANAL CHIM ACTA, V166, P111
[9]   GRAPHITE PASTE-BASED ENZYMATIC GLUCOSE ELECTRODE FOR FLOW-INJECTION ANALYSIS [J].
MATUSZEWSKI, W ;
TROJANOWICZ, M .
ANALYST, 1988, 113 (05) :735-738
[10]   AMPEROMETRIC ENZYME ELECTRODE SYSTEM FOR THE FLOW-INJECTION ANALYSIS OF GLUCOSE [J].
MOODY, GJ ;
SANGHERA, GS ;
THOMAS, JDR .
ANALYST, 1986, 111 (06) :605-609