MEASUREMENT OF SODIUM-HYDROXIDE CONCENTRATION WITH A RENEWABLE REAGENT-BASED FIBEROPTIC SENSOR

被引:45
作者
BERMAN, RJ [1 ]
CHRISTIAN, GD [1 ]
BURGESS, LW [1 ]
机构
[1] UNIV WASHINGTON, CTR PROC ANALYT CHEM, DEPT CHEM, BG-10, SEATTLE, WA 98195 USA
关键词
D O I
10.1021/ac00218a003
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fiber-optic caustic sensor Is described In which a weakly buffered acid-base Indicator reagent (bromothymol blue) is continuously delivered to the sensor tip which consists of a length of Nation hollow fiber membrane, an analyte exchange region, and an optical cell. In this study, caustic concentrations of 1.5-23.0% (w/w) (0.38-7.2 M) were measured, with a relative standard deviation of 1.6%. Sensor performance was studied at several reagent compositions, at various reagent delivery rates, In the presence of brine, and with respect to probe-to-probe reproducibility and long-term use. This technique, of controlled reagent phase renewal, successfully addresses several of the limitations of previous reagent-based fiber-optic sensors. In addition, many new opportunities are created when considering the flexibility of such sensors, such as the ability to use multiple reagents and Irreversible chemistries. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:2066 / 2071
页数:6
相关论文
共 12 条
[1]  
BERMAN RJ, 1990, P SOC PHOTO-OPT INS, V1172, P206, DOI 10.1117/12.963190
[2]  
CHRISTIAN GD, 1980, ANAL CHEM, P304
[3]  
CLEVITT KJ, 1986, PROCESS ANAL TECHNOL, P506
[4]  
IMATO T, 1985, ANAL SCI, V1, P481
[5]   PRESSURIZED MEMBRANE INDICATOR SYSTEM FOR FLUOROGENIC-BASED FIBER-OPTIC CHEMICAL SENSORS [J].
INMAN, SM ;
STROMVALL, EJ ;
LIEBERMAN, SH .
ANALYTICA CHIMICA ACTA, 1989, 217 (02) :249-262
[6]   FIBER-OPTIC SENSORS BASED ON REAGENT DELIVERY WITH CONTROLLED-RELEASE POLYMERS [J].
LUO, SF ;
WALT, DR .
ANALYTICAL CHEMISTRY, 1989, 61 (02) :174-177
[7]   THERMOMETRIC ION-EXCHANGE - AN APPROACH FOR MEASUREMENT OF ACID OR BASE CONCENTRATION ABOVE 1-N [J].
MILLER, TE ;
SMALL, H .
ANALYTICAL CHEMISTRY, 1985, 57 (01) :262-265
[8]  
MILLER TE, 1980, Patent No. 4199323
[9]   MEASUREMENT OF CAUSTIC AND CAUSTIC BRINE SOLUTIONS BY SPECTROSCOPIC DETECTION OF THE HYDROXIDE ION IN THE NEAR-INFRARED REGION, 700-1150 NM [J].
PHELAN, MK ;
BARLOW, CH ;
KELLY, JJ ;
JINGUJI, TM ;
CALLIS, JB .
ANALYTICAL CHEMISTRY, 1989, 61 (13) :1419-1424
[10]  
SEITZ WR, 1988, CRC CRIT R ANAL CHEM, V57, P173