BALLOON ATRIAL SEPTOSTOMY UNDER 2-DIMENSIONAL ECHOCARDIOGRAPHIC CONTROL

被引:11
作者
BEITZKE, A [1 ]
STEIN, JI [1 ]
SUPPAN, C [1 ]
机构
[1] GRAZ UNIV,CHILDRENS HOSP,DEPT PEDIAT CARDIOL,A-8036 GRAZ,AUSTRIA
关键词
BALLOON ATRIAL SEPTOSTOMY; 2-DIMENSIONAL ECHOCARDIOGRAPHY; TRANSPOSITION OF THE GREAT ARTERIES; EUSTACHIAN VALVE;
D O I
10.1016/0167-5273(91)90121-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We report our experience with balloon atrial septostomy under two-dimensional echocardiographic guidance only in 44 newborn patients (39 patients with transposition of the great arteries) who needed creation of an interatrial communication. Using standard echocardiographic projections septostomy with creation of large atrial septal defects was easy in 41 newborns. In 3 patients insertion of a large Eustachian valve at the interatrial septum prevented adequate pull-throughs with larger balloon sizes. No serious complications due to the procedure were noted. Balloon atrial septostomy under two-dimensional echocardiographic control is a quick, effective and safe method which can be performed on the intensive care unit.
引用
收藏
页码:33 / 42
页数:10
相关论文
共 13 条
  • [1] AGUILAR MB, 1984, PEDIATR CARDIOL, V5, P248
  • [2] ALLAN LD, 1982, BRIT HEART J, V47, P41
  • [3] BAKER EJ, 1984, BRIT HEART J, V51, P377
  • [4] BEITZKE A, 1985, Z KARDIOL, V74, P353
  • [5] BEITZKE A, 1983, BRIT HEART J, V49, P341
  • [6] PROSPECTIVE DIAGNOSIS OF D-TRANSPOSITION OF THE GREAT-ARTERIES IN NEONATES BY SUBXIPHOID, 2-DIMENSIONAL ECHOCARDIOGRAPHY
    BIERMAN, FZ
    WILLIAMS, RG
    [J]. CIRCULATION, 1979, 60 (07) : 1496 - 1502
  • [7] BEDSIDE BALLOON ATRIAL SEPTOSTOMY USING ECHOCARDIOGRAPHIC MONITORING
    BULLABOY, CA
    JENNINGS, RB
    JOHNSON, DH
    FULCHER, CW
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (07) : 971 - 971
  • [8] ACCURACY OF 2-DIMENSIONAL ECHOCARDIOGRAPHY IN THE DIAGNOSIS OF CONGENITAL HEART-DISEASE
    GUTGESELL, HP
    HUHTA, JC
    LATSON, LA
    HUFFINES, D
    MCNAMARA, DG
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1985, 55 (05) : 514 - 518
  • [9] ECHOCARDIOGRAPHIC ANATOMY OF THE EUSTACHIAN VALVE
    LIMACHER, MC
    GUTGESELL, HP
    VICK, GW
    COHEN, MH
    HUHTA, JH
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (04) : 363 - 365
  • [10] BALLOON AND BLADE ATRIAL SEPTOSTOMY FACILITATED BY TWO-DIMENSIONAL ECHOCARDIOGRAPHY
    LIN, AE
    DISESSA, TG
    WILLIAMS, RG
    LEIGHTON, J
    GROSS, K
    WONG, AL
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (04) : 273 - 277