ANALYTICAL ARTIFACTS IN HEMATOCRIT MEASUREMENTS BY WHOLE-BLOOD CHEMISTRY ANALYZERS

被引:0
作者
STOTT, RAW
HORTIN, GL
WILHITE, TR
MILLER, SB
SMITH, CH
LANDT, M
机构
[1] WASHINGTON UNIV,SCH MED,DEPT PATHOL,ST LOUIS,MO 63110
[2] WASHINGTON UNIV,SCH MED,EDWARD MALLINCKRODT DEPT PEDIAT,ST LOUIS,MO 63110
[3] WASHINGTON UNIV,SCH MED,DEPT MED,ST LOUIS,MO 63110
[4] ST LOUIS CHILDRENS HOSP,ST LOUIS,MO 63110
关键词
CONDUCTIVITY; NEAR-PATIENT TESTING; OSMOLALITY; INTERMETHOD COMPARISON; ANALYTICAL ERROR;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Compact analyzers suited to near-patient testing estimate hematocrit by measuring the conductivity of undiluted blood. We evaluated the accuracy of hematocrit determination of one such analyzer (Instrumentation Laboratory BGE Analyzer) against an automated cell counter (EPC) and packed cell volume (PCV) microhematocrit. When specimens (n = 34) from outpatient and ward patients were analyzed with all three methods, the BGE analyzer correlated well with both EPC and PCV hematocrit determinations (BGE = 1.00 PCV + 0.3%, S-y/x = 1.6%; BGE = 1.04 EPC + 0.4%, S-y/x = 2.0%), suggesting that all three methods are similar in performance for most patients. However, a patient with increased plasma osmolality showed significant decreases in BGE and PCV hematocrits relative to the EPC method. The differences in hematocrit measurements could be reproduced by adding solutes to blood in vitro or by modifying the plasma osmolality of rats in vivo. Samples from patients undergoing cardiac surgery, whose blood had large changes in protein concentration, showed discrepancies between hematocrits by conductivity and other methods; similar effects could be produced by changes in protein concentration or in vitro addition of polyethylene glycol. We conclude that conductivity measurements provide accurate hematocrit results for physiologically normal subjects but not for some intensive-care and surgical patients.
引用
收藏
页码:306 / 311
页数:6
相关论文
共 16 条
[1]   EVALUATION OF A BLOOD-GAS AND CHEMISTRY MONITOR FOR USE DURING SURGERY [J].
BASHEIN, G ;
GREYDANUS, WK ;
KENNY, MA .
ANESTHESIOLOGY, 1989, 70 (01) :123-127
[2]  
BEAUTYMAN W, 1974, LANCET, V2, P905
[4]  
CROW DR, 1974, CONDUCTANCE ELECTROL, P52
[5]  
ENGLAND JM, 1991, PRACTICAL LABORATORY, P109
[6]  
FLEISHER M, 1989, CLIN CHEM, V35, P1532
[7]  
FRICKE H, 1925, PHYS REV 2, V24, P575
[8]   SPECIFIC RESISTANCE OF BLOOD AT BODY-TEMPERATURE [J].
GEDDES, LA ;
SADLER, C .
MEDICAL & BIOLOGICAL ENGINEERING, 1973, 11 (03) :336-339
[9]  
Gotch F, 1991, ASAIO Trans, V37, pM138
[10]  
Maxwell J. C., 1904, TREATISE ELECT MAGNE, Vsecond, P435