A SENSITIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF 6-MERCAPTOPURINE IN PLASMA USING PRECOLUMN DERIVATIZATION AND FLUORESCENCE DETECTION

被引:14
|
作者
WARREN, DJ
SLORDAL, L
机构
[1] Department of Clinical Pharmacology, The Norwegian Radium Hospital, Oslo
关键词
6-MERCAPTOPURINE; HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY; FLUORESCENCE DETECTION;
D O I
10.1097/00007691-199302000-00004
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A sensitive high-performance liquid chromatographic (HPLC) method for measuring plasma concentrations of 6-mercaptopurine (6-MP) is described. After protein precipitation with 5-sulfosalicylic acid, samples are subjected to precolumn derivatization using the thiol-reactive fluorophore monobromobimane (mBrB). The drug-mBrB adduct is then resolved by isocratic elution from a C18 reversed-phase support and quantified by fluorescence detection. Recovery of 6-MP after protein precipitation was consistently >85% and the drug-mBrB adduct was found to be stable for at least 2 weeks at room temperature. With plasma samples containing 30 nM 6-MP, the assay displayed within-run (n = 6) and between-day (n = 6) coefficients of variation of 2.2 and 10.6%, respectively. The limit of detection for 6-MP in plasma was 3 nM (500 pg/ml) and the standard curve was linear up to 3 muM. Using this method, we have observed that 6-MP is stable in heparinized whole blood for at least 24 h provided samples are maintained on ice. Since this method requires few manipulations during sample preparation and is readily adaptable to automated techniques, it may prove useful in the routine clinical laboratory setting.
引用
收藏
页码:25 / 30
页数:6
相关论文
共 50 条
  • [21] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF PEPTIDES IN BIOLOGICAL-FLUIDS BY AUTOMATED PRECOLUMN FLUORESCENCE DERIVATIZATION WITH FLUORESCAMINE
    BOPPANA, VK
    MILLERSTEIN, C
    POLITOWSKI, JF
    RHODES, GR
    JOURNAL OF CHROMATOGRAPHY, 1991, 548 (1-2): : 319 - 327
  • [22] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR MONITORING LEUPEPTIN IN MOUSE SERUM AND MUSCLE BY PRECOLUMN FLUORESCENCE DERIVATIZATION WITH BENZOIN
    KAI, M
    MIURA, T
    ISHIDA, J
    OHKURA, Y
    JOURNAL OF CHROMATOGRAPHY, 1985, 345 (02): : 259 - 265
  • [23] LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF CALCIUM CYANAMIDE USING PRECOLUMN DERIVATIZATION
    CHEN, S
    OCAMPO, AP
    KUCERA, PJ
    JOURNAL OF CHROMATOGRAPHY, 1991, 558 (01): : 141 - 146
  • [24] LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF CALCIUM CYANAMIDE USING PRECOLUMN DERIVATIZATION
    CHEN, S
    OCAMPO, A
    KUCERA, P
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1991, 202 : 113 - ANYL
  • [25] ISOLATION OF 6-MERCAPTOPURINE IN HUMAN PLASMA BY ALUMINUM ION COMPLEXATION FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS
    LIN, KT
    VARIN, F
    RIVARD, GE
    LECLERC, JM
    JOURNAL OF CHROMATOGRAPHY, 1991, 536 (1-2): : 349 - 355
  • [26] LIQUID-CHROMATOGRAPHIC DETERMINATION OF SISOMICIN IN PLASMA, WITH FLUOROMETRIC PRECOLUMN DERIVATIZATION
    KOSHIDE, K
    TAWA, R
    HIROSE, S
    FUJIMOTO, T
    CLINICAL CHEMISTRY, 1985, 31 (11) : 1921 - 1922
  • [27] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF PENBUTOLOL ENANTIOMERS IN PLASMA WITH FLUORESCENCE DETECTION
    GOTO, J
    SHAO, G
    ITO, M
    KURIKI, T
    NAMBARA, T
    ANALYTICAL SCIENCES, 1991, 7 (05) : 723 - 726
  • [28] A RAPID AND SENSITIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY FOR 6-MERCAPTOPURINE METABOLITES IN RED BLOOD-CELLS
    LAVI, LE
    HOLCENBERG, JS
    ANALYTICAL BIOCHEMISTRY, 1985, 144 (02) : 514 - 521
  • [29] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF ETOPOSIDE IN PLASMA USING ELECTROCHEMICAL DETECTION
    DUNCAN, GF
    FARMEN, RH
    MOVAHHED, HS
    PITTMAN, KA
    JOURNAL OF CHROMATOGRAPHY, 1986, 380 (02): : 357 - 365
  • [30] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF URINARY CATECHOLAMINES BY DIRECT PRECOLUMN FLUORESCENCE DERIVATIZATION WITH 1,2-DIPHENYLETHYLENEDIAMINE
    NOHTA, H
    MITSUI, A
    OHKURA, Y
    JOURNAL OF CHROMATOGRAPHY, 1986, 380 (01): : 229 - 231