EVALUATION OF SERUM DIGOXIN AND SERUM CORTISOL BY RADIOIMMUNOASSAY WITH THE CONCEPT-4

被引:2
作者
WALSH, PR
ABARCA, SD
WANG, MC
机构
[1] Department of Chemistry, Laboratory. Procedures East, Inc. Subsidiary, the Upjohn Company, King of Prussia
关键词
D O I
10.1016/S0009-9120(79)80136-8
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The solid-phase radioimmunoassay procedures for serum digoxin and cortisol with the Concept-4 were evaluated for specificity, correlation with other radioimmunoassay procedures, parallelism, recovery, carryover, and precision.1.The detection limits of the digoxin and cortisol assays were 10 pg/tube and 40 pg/tube, respectively. These detections correspond to serum concentrations of 0.25 μg/liter and 20 μg/liter, respectively, for digoxin and cortisol.2.Specificity of the digoxin assay was characterized by a 18.3% cross reaction with digitoxin. The anti-cortisol with 11-deoxycortisol. cross reaction with 11-deoxycortisol.3.The values obtained with the Concept-4 for both procedures demonstrated acceptable correlations with other existing RIA procedures.4.Dilution studies conducted to determine the parallelism of the 2 assays indicated that both assays demonstrated excellent proportionality between the volume of specimen and the quantity of analyte measured.5.Analytical recoveries of digoxin and cortisol added to the individual assays average 95% and 101%, respectively.6.Carryover studies indicated a routine sample-to-sample interaction of < 0.05% for both assays.7.The digoxin assay routinely demonstrated intra- and inter-assay variations of < 7.4% and < 10.0%, respectively. The precision data for cortisol indicated intra- and inter-assay variations of < 6.7% and < 9.1%, respectively. © 1979.
引用
收藏
页码:118 / 122
页数:5
相关论文
共 50 条
  • [31] Local preparation and evaluation of liquid phase radioimmunoassay for determination of human serum cortisol
    Sallam, Kh. M.
    Zein, N.
    El-Bayoumy, A. S. A.
    Ebeid, N. H.
    Michael, E.
    Elfarargy, Ah. F.
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 313 (01) : 161 - 168
  • [32] A RAPID SOLID-PHASE RADIOIMMUNOASSAY (DIGOXIN RIABEADTM) FOR THE DETERMINATION OF DIGOXIN IN SERUM
    HO, TT
    [J]. CLINICAL CHEMISTRY, 1981, 27 (06) : 1088 - 1088
  • [33] DIRECT RADIOIMMUNOASSAY OF CORTISOL IN SERUM BY ENZYMATIC DENATURATION OF SERUM BINDING-PROTEINS
    PAINTER, K
    HASLER, M
    NISWENDER, G
    [J]. CLINICAL CHEMISTRY, 1975, 21 (07) : 1023 - 1023
  • [34] EVALUATION OF SERUM THYROXINE BY RADIOIMMUNOASSAY
    BECKERS, C
    CORNETTE, C
    THALASSO, M
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1973, 14 (06) : 317 - 320
  • [35] DETERMINATION OF DIGOXIN IN SERUM - COMPARISON OF RADIOIMMUNOASSAY AND A HETEROGENEOUS ENZYME IMMUNOASSAY
    BORNER, K
    RIETBROCK, N
    [J]. JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1978, 16 (06): : 335 - 342
  • [36] ON THE COMPARISON OF SERUM CORTISOL BY DIRECT RADIOIMMUNOASSAY AND FLUOROMETRIC METHOD
    KANLUAN, T
    PLOYBUTR, S
    VICHAYANRAT, A
    [J]. JOURNAL OF THE MEDICAL ASSOCIATION OF THAILAND, 1982, 65 (05): : 263 - 270
  • [37] PROBLEMS IN MEASUREMENT OF SERUM DIGOXIN BY COMMERCIALLY AVAILABLE RADIOIMMUNOASSAY KITS
    KUBASIK, NP
    BRODY, BB
    BAROLD, SS
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1975, 36 (07) : 975 - 977
  • [38] EFFECT OF PALLADIUM ON THE QUANTITATION OF SERUM DIGOXIN USING RADIOIMMUNOASSAY TECHNIQUE
    LIU, TZ
    [J]. CLINICAL CHEMISTRY, 1979, 25 (06) : 1135 - 1135
  • [39] DETERMINATION OF SERUM DIGOXIN WITH SIMPLE H-3 RADIOIMMUNOASSAY
    KLEESIEK, K
    TARKKANEN, P
    VANWERSCH, J
    [J]. JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1977, 15 (03): : 166 - 166
  • [40] METHODOLOGICAL CAUSES OF DISCREPANCIES IN RADIOIMMUNOASSAY FOR DIGOXIN IN HUMAN-SERUM
    HOLTZMAN, JL
    SHAFER, RB
    ERICKSON, RR
    [J]. CLINICAL CHEMISTRY, 1974, 20 (09) : 1194 - 1198