EVALUATION OF SERUM DIGOXIN AND SERUM CORTISOL BY RADIOIMMUNOASSAY WITH THE CONCEPT-4

被引:2
作者
WALSH, PR
ABARCA, SD
WANG, MC
机构
[1] Department of Chemistry, Laboratory. Procedures East, Inc. Subsidiary, the Upjohn Company, King of Prussia
关键词
D O I
10.1016/S0009-9120(79)80136-8
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The solid-phase radioimmunoassay procedures for serum digoxin and cortisol with the Concept-4 were evaluated for specificity, correlation with other radioimmunoassay procedures, parallelism, recovery, carryover, and precision.1.The detection limits of the digoxin and cortisol assays were 10 pg/tube and 40 pg/tube, respectively. These detections correspond to serum concentrations of 0.25 μg/liter and 20 μg/liter, respectively, for digoxin and cortisol.2.Specificity of the digoxin assay was characterized by a 18.3% cross reaction with digitoxin. The anti-cortisol with 11-deoxycortisol. cross reaction with 11-deoxycortisol.3.The values obtained with the Concept-4 for both procedures demonstrated acceptable correlations with other existing RIA procedures.4.Dilution studies conducted to determine the parallelism of the 2 assays indicated that both assays demonstrated excellent proportionality between the volume of specimen and the quantity of analyte measured.5.Analytical recoveries of digoxin and cortisol added to the individual assays average 95% and 101%, respectively.6.Carryover studies indicated a routine sample-to-sample interaction of < 0.05% for both assays.7.The digoxin assay routinely demonstrated intra- and inter-assay variations of < 7.4% and < 10.0%, respectively. The precision data for cortisol indicated intra- and inter-assay variations of < 6.7% and < 9.1%, respectively. © 1979.
引用
收藏
页码:118 / 122
页数:5
相关论文
共 50 条
[21]   SIMPLE AND ECONOMICAL METHOD FOR RADIOIMMUNOASSAY OF CORTISOL IN SERUM [J].
MORRIS, R .
ANNALS OF CLINICAL BIOCHEMISTRY, 1978, 15 (MAY) :178-183
[22]   RADIOIMMUNOASSAY OF SERUM DIGOXIN AND DIGITOXIN IN TOXICITY AND MALABSORPTION STATES [J].
SMITH, TW ;
HABER, E .
CIRCULATION, 1969, 40 (4S3) :I188-&
[23]   REPRODUCIBILITY OF STANDARD PREPARATION IN DIGOXIN RADIOIMMUNOASSAY IN PLASMA AND SERUM [J].
MOLIN, L ;
BERGDAHL, B ;
DAHLSTROM, G ;
LEANDER, E .
THERAPEUTIC DRUG MONITORING, 1983, 5 (03) :355-357
[24]   DETERMINATION OF TERAPEUTIC AND TOXIC SERUM DIGOXIN CONCENTRATIONS BY RADIOIMMUNOASSAY [J].
SMITH, TW ;
BUTLER, VP ;
HABER, E .
NEW ENGLAND JOURNAL OF MEDICINE, 1969, 281 (22) :1212-+
[25]   RADIOIMMUNOASSAY FOR SERUM DIGOXIN, CLINICALLY IMPORTANT CROSS REACTIONS [J].
REISERT, PM ;
KISTNER, M .
FORTSCHRITTE DER MEDIZIN, 1977, 95 (03) :167-171
[26]   APPLICATION OF RADIOIMMUNOASSAY TECHNIQUES FOR DETERMINING SERUM LEVELS OF DIGOXIN [J].
CHAMBERLAIN, DA ;
WHITE, RJ .
CLINICAL SCIENCE, 1969, 37 (02) :572-+
[27]   RESULTS COMPARED FOR RADIOIMMUNOASSAY OF DIGOXIN USING A SOLID-PHASE MANUAL PROCEDURE AND A TOTALLY AUTOMATED-METHOD USING CONCEPT-4 [J].
WARKENTIN, DL ;
VANLENTE, F ;
OHNO, T .
CLINICAL CHEMISTRY, 1977, 23 (06) :1124-1124
[28]   RADIOIMMUNOASSAY OF SALIVARY DIGOXIN BY SIMPLE ADAPTATION OF A KIT METHOD FOR SERUM DIGOXIN - SALIVA-SERUM RATIO AND CORRELATION [J].
MAHMOD, S ;
SMITH, DS ;
LANDON, J .
THERAPEUTIC DRUG MONITORING, 1987, 9 (01) :91-96
[29]   STUDIES OF MONOCLONAL AND POLYCLONAL ANTI-DIGOXIN ANTIBODIES FOR SERUM DIGOXIN RADIOIMMUNOASSAY [J].
BANG, BE ;
HURME, M ;
JUNTUNEN, K ;
MAKELA, O .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1981, 41 (01) :75-78
[30]   Local preparation and evaluation of liquid phase radioimmunoassay for determination of human serum cortisol [J].
Kh. M. Sallam ;
N. Zein ;
A. S. A. El-Bayoumy ;
N. H. Ebeid ;
E. Michael ;
Ah. F. Elfarargy .
Journal of Radioanalytical and Nuclear Chemistry, 2017, 313 :161-168