TESTS OF EEG LOCALIZATION ACCURACY USING IMPLANTED SOURCES IN THE HUMAN BRAIN

被引:102
作者
CUFFIN, BN
COHEN, D
YUNOKUCHI, K
MANIEWSKI, R
PURCELL, C
COSGROVE, GR
IVES, J
KENNEDY, J
SCHOMER, D
机构
[1] BETH ISRAEL HOSP,DEPT SURG,BOSTON,MA 02215
[2] KAGOSHIMA UNIV,KAGOSHIMA 890,JAPAN
[3] POLISH ACAD SCI,WARSAW 42,POLAND
[4] BETH ISRAEL HOSP,DEPT NEUROL,BOSTON,MA 02215
[5] HITCHCOCK CLIN,MANCHESTER,NH
关键词
D O I
10.1002/ana.410290204
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The accuracy with which electrical sources in the brain can be localized by using electroencephalograms measured on the scalp is not well known. In this study, tests of localization accuracy were performed by using implanted dipolar sources in the human brain. These dipoles are created by passing a weak (subthreshold) current through intracerebral electrodes implanted in the brains of epileptic patients for seizure monitoring. The locations of these dipoles are accurately known from roentgenographs. First, 16 electroencephalograms. Finally, the locations of the apparent sources were compared with the known locations of the implanted dipoles to determine localization error. The average localization error for a total of 28 dipoles in 3 subjects was 1.1 cm. These results indicate that good localization accuracy for focal sources in the brain can be provided by scalp electroencephalograms.
引用
收藏
页码:132 / 138
页数:7
相关论文
共 15 条
[1]   LOCATION OF SOURCES OF EVOKED SCALP POTENTIALS - CORRECTIONS FOR SKULL AND SCALP THICKNESSES [J].
ARY, JP ;
KLEIN, SA ;
FENDER, DH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (06) :447-452
[2]   MEG VERSUS EEG LOCALIZATION TEST USING IMPLANTED SOURCES IN THE HUMAN BRAIN [J].
COHEN, D ;
CUFFIN, BN ;
YUNOKUCHI, K ;
MANIEWSKI, R ;
PURCELL, C ;
COSGROVE, GR ;
IVES, J ;
KENNEDY, JG ;
SCHOMER, DL .
ANNALS OF NEUROLOGY, 1990, 28 (06) :811-817
[4]   EFFECTS OF HEAD SHAPE ON EEGS AND MEGS [J].
CUFFIN, BN .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1990, 37 (01) :44-52
[6]  
Darcey T M, 1980, Prog Brain Res, V54, P128, DOI 10.1016/S0079-6123(08)61617-9
[7]   ACCURACY OF DIPOLE LOCALIZATION WITH A SPHERICAL HOMOGENEOUS MODEL [J].
GAUMOND, RP ;
LIN, JH ;
GESELOWITZ, DB .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1983, 30 (01) :29-34
[8]   ELECTRIC-DIPOLE TRACING IN THE BRAIN BY MEANS OF THE BOUNDARY ELEMENT METHOD AND ITS ACCURACY [J].
HE, B ;
MUSHA, T ;
OKAMOTO, Y ;
HOMMA, S ;
NAKAJIMA, Y ;
SATO, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1987, 34 (06) :406-414
[9]   EVALUATION OF METHODS FOR 3-DIMENSIONAL LOCALIZATION OF ELECTRICAL SOURCES IN HUMAN-BRAIN [J].
KAVANAGH, RN ;
DARCEY, TM ;
LEHMANN, D ;
FENDER, DH .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1978, 25 (05) :421-429
[10]  
OKADA Y, 1985, BIOMAGNETISM APPL TH, P266