PREVENTION OF CATHETER-RELATED SEPSIS DURING PARENTERAL-NUTRITION - EFFECT OF A NEW CONNECTION DEVICE

被引:30
作者
INOUE, Y
NEZU, R
MATSUDA, H
FUJII, M
NAKAI, S
WASA, M
TAKAGI, Y
OKADA, A
机构
[1] OSAKA UNIV,SCH MED,DEPT PEDIAT SURG,1-1-50 FUKUSHIMA,OSAKA 553,JAPAN
[2] OSAKA UNIV,SCH MED,DEPT SURG 1,OSAKA 553,JAPAN
关键词
D O I
10.1177/0148607192016006581
中图分类号
R15 [营养卫生、食品卫生]; TS201 [基础科学];
学科分类号
100403 ;
摘要
A prospective study was carried out to determine the clinical effect of a newly devised catheter connection method (I system) and piggyback access system. Previous studies have demonstrated that the I system avoided bacterial contamination in vitro during tubing change that Luer-Lock connectors did not. The purpose of this study was to investigate the ability of this device coupled with a new closed-system piggyback technique for multipurpose access to reduce catheter-related sepsis in clinical practice. Two hundred and thirty patients receiving total parenteral nutrition were divided into two groups. Group I (n = 106) used the I system connector and group L (n = 124) used a Luer-Lock connector. Catheters in both groups were used for multipurpose access for infusion and blood sampling. In group L, a three-way stopcock and/or piggyback system was used for multiple access. In group I, a newly designed closed-system piggyback was used. The incidence of catheter-related sepsis was significantly lower in group I (1.89%/catheter) than in group L (12.10%/catheter) (p < .01, chi2 analysis), and the average duration of use of each catheter was significantly longer in group I than in Group L (p < .01 by generalized Wilcoxon test). The results of this clinical study suggest that the newly designed connection method and piggyback access system are able to reduce catheter-related sepsis.
引用
收藏
页码:581 / 585
页数:5
相关论文
共 18 条
[1]  
BOZZETTI F, 1982, SURGERY, V91, P383
[2]   IMPLICATION OF A DISTANT SEPTIC FOCUS IN PARENTERAL-NUTRITION CATHETER COLONIZATION [J].
CHUANG, JH ;
CHUANG, SF .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1991, 15 (02) :173-175
[3]   AN OUTBREAK OF STAPHYLOCOCCUS-EPIDERMIDIS SEPTICEMIA [J].
DEITEL, M ;
KRAJDEN, S ;
SALDANHA, CF ;
GREGORY, WD ;
FUKSA, M ;
CANTWELL, E .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1983, 7 (06) :569-572
[4]  
DRYDEN GE, 1979, ANESTH ANALG, V58, P141
[5]   TOTAL PARENTERAL-NUTRITION CATHETER SEPSIS - IMPACT OF THE TEAM-APPROACH [J].
FAUBION, WC ;
WESLEY, JR ;
KHALIDI, N ;
SILVA, J .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1986, 10 (06) :642-645
[6]  
FORSE RA, 1979, SURGERY, V86, P507
[7]   INVITRO CONTAMINATION OF PIGGYBACK HEPARIN LOCK ASSEMBLIES - PREVENTION OF CONTAMINATION WITH A CLOSED, POSITIVE LOCKING DEVICE (CLICK-LOCK) [J].
GIBILISCO, PA ;
LOPEZ, GA ;
APPLEMAN, MD ;
MILLBURG, WJ ;
KUSHELL, DL ;
ELOWITZ, L ;
JOHNSON, PA ;
ZENOWICH, D .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1986, 10 (04) :431-434
[8]  
HOLM I, 1975, ACTA CHIR SCAND, V141, P173
[9]   MICROBIAL-CONTAMINATION OF INTRAVENOUS INFUSIONS DURING CLINICAL USE [J].
HOLMES, CJ ;
ALLWOOD, MC .
JOURNAL OF APPLIED BACTERIOLOGY, 1979, 46 (02) :247-267
[10]   EXPERIMENTAL-STUDY OF HUB CONTAMINATION - EFFECT OF A NEW CONNECTION DEVICE - THE I SYSTEM [J].
INOUE, Y ;
NEZU, R ;
MATSUDA, H ;
FUJII, M ;
NAKAI, S ;
WASA, M ;
TAKAGI, Y ;
OKADA, A .
JOURNAL OF PARENTERAL AND ENTERAL NUTRITION, 1992, 16 (02) :178-180