Treatment planning evaluation of sliding window and multiple static segments technique in intensity modulated radiotherapy

被引:17
作者
Iqbal, Khalid [1 ,2 ,3 ]
Isa, Muhammad [1 ]
Buzdar, Saeed Ahmad [1 ]
Gifford, Kent Aallen [3 ]
Afzal, Muhammad. [1 ]
机构
[1] Islamia Univ Bahawalpur, Dept Phys, Bahawalpur, Pakistan
[2] Shaukat Khanum Canc Hosp & Res Ctr, Dept Radiat Oncol, Lahore, Pakistan
[3] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
Conformity index; Homogeneity index; Uniformity index; Organs-at-risk; Intensity modulated radiation therapy;
D O I
10.1016/j.rpor.2012.10.003
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: The demand of improved dose conformity of the tumor has been increased in radiation therapy with the advent of recent imaging facilities and efficient computer technologies. Aim: We compared the intensity modulated radiotherapy (IMRT) plans delivered with the sliding window (SW IMRT) and step and shoot (SS IMRT) techniques. Materials and methods: Thirteen patients were planned on 15 MV X-ray for five, seven, nine and thirteen beams direction making the dose constraints analogous. Eclipse treatment planning system with Helios inverse planning software, and Linear Accelerator Varian 2100 C/D with 120 multileaf collimators (MLCs) were used. Gamma analysis was applied to the data acquired with the MapCheck 2 (TM) for different beam directions plan in the sliding window and step and shoot technique to meet the 95% pass criteria at 3%/3 mm. The plans were scrutinized using D-mean, D-max, D1%, D95%, dose uniformity index (UI), dose conformity index (CI), dose homogeneity index (HI) and monitor units (MUs). Results: Our data show comparable coverage of the planning target volume (PTV) for both the sliding window and step and shoot techniques. The volume of PTV receiving the prescription dose was 99.8 +/- 0.05% and the volume of PTV receiving the maximum dose was 107.6 +/- 2.5% in both techniques. Bladder and rectum maximum mean doses for the sliding window and step and shoot plans were 38.1 +/- 2.6% and 42.9 +/- 10.7%. Homogeneity index (HI) for both techniques was 0.12 +/- 0.02 and 0.13 +/- 0.02, uniformity index (UI) was 1.07 +/- 0.02 and 108 +/- 0.01 and conformity index at 98% isodose (CI 98%) was 0.96 +/- 0.005 and 0.96 +/- 0.005 for the sliding window and step and shoot techniques, respectively, and MUs were 10 +/- 12% lower in the step and shoot compared to the sliding window technique. Conclusion: All these factors indicate that coverage for PTV was nearly identical but dose to organs-at-risk (OARs) was lower in the step and shoot technique. (C) 2012 Greater Poland Cancer Centre. Published by Elsevier Urban & Partner Sp. z o.o. All rights reserved.
引用
收藏
页码:101 / 106
页数:6
相关论文
共 19 条
[1]  
Alaei Parham, 2004, Med Dosim, V29, P1, DOI 10.1016/j.meddos.2003.10.002
[2]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[3]  
Blanco AI, 2002, INTENSITY MODULATED, V3
[4]   Delivery of intensity-modulated radiation therapy with a conventional multileaf collimator: Comparison of dynamic and segmental methods [J].
Chui, CS ;
Chan, MF ;
Yorke, E ;
Spirou, S ;
Ling, CC .
MEDICAL PHYSICS, 2001, 28 (12) :2441-2449
[5]   Radiation-induced second cancers: The impact of 3D-CRT and IMRT [J].
Hall, EJ ;
Wuu, CS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :83-88
[6]   Intensity-modulated radiation therapy, protons, and the risk of second cancers [J].
Hall, Eric J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 65 (01) :1-7
[7]  
Hande BA, 2011, REP PRACT ONCOL RADI, V16, P102
[8]   The calculated risk of fatal secondary malignancies from intensity-modulated radiation therapy [J].
Kry, SF ;
Salehpour, M ;
Followill, DS ;
Stovall, M ;
Kuban, DA ;
White, RA ;
Rosen, II .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1195-1203
[9]  
Malhotra Harish K, 2005, J APPL CLIN MED PHYS, V6
[10]   The impact of fluctuations in intensity patterns on the number of monitor units and the quality and accuracy of intensity modulated radiotherapy [J].
Mohan, R ;
Arnfield, M ;
Tong, SD ;
Wu, QW ;
Siebers, J .
MEDICAL PHYSICS, 2000, 27 (06) :1226-1237