Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging system

被引:24
作者
Sharma, Reena [1 ]
Sharma, Sunil Dutt [1 ]
Pawar, Shramika [1 ]
Chaubey, Ajay [2 ]
Kantharia, S. [2 ]
Babu, D. A. R. [1 ]
机构
[1] CT & CRS, Radiol Phys & Advisory Div, Bombay, Maharashtra, India
[2] Bhabha Atom Res Ctr Hosp, Bombay, Maharashtra, India
关键词
Patient dose; entrance skin dose; diagnostic reference level; computed radiography; CR imaging system; X-ray machine; X-ray imaging;
D O I
10.4103/0971-6203.152244
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The screen-film system is replaced by computed radiography system for recording the images of the patients during X-ray radiography examinations. The change in imaging system requires the re-establishment of the institutional diagnostic reference levels (DRLs) for different types of X-ray examinations conducted at the hospital. For this purpose, patient specific parameters [age, height, weight, body mass index (BMI), object to image distance (OID)] and machine specific parameters (kVp, mAs, distance and field sizes) of 1875 patients during 21 different types of X-ray examinations were recorded for estimating the entrance skin dose (ESD). The ESD for each of these patients were estimated using measured X-ray beam output and the standard value of the back scatter factor. Five number summary was calculated for all the data for their presentation in the Box-Whisker plot, which provides the statistical distribution of the data. The data collected indicates that majorly performed examinations are cervical spine AP, Chest PA and Knee Lat with percentage contributions of 16.05, 16 and 8.27% respectively. The lowest contribution comes from Hip Lat which is about 1.01%. The ratio of measured ESD (maximum to minimum) for these examinations is found to be highest for the cervical spine AP with a value of 50 followed by Thoracic spine AP of 32.36. The ESD ratio for Chest PA, Knee Lat and Lumbar Spine AP are 30.75, 30.4 and 30.2 respectively. The lowest ESD ratio is for Hip Lat which is 2.68. The third quartile values of ESDs are established as the institutional DRLs. The ESD values obtained for 21 different X-ray projections are either comparable or lesser than the reported national/international values.
引用
收藏
页码:29 / 37
页数:9
相关论文
共 50 条
  • [31] X-ray Fluorescence Computed Tomography (XFCT) Imaging with a Superfine Pencil Beam X-ray Source
    Romero, Ignacio O.
    Fang, Yile
    Lun, Michael
    Li, Changqing
    PHOTONICS, 2021, 8 (07)
  • [32] Dose Correction in Medical X-ray Imaging in Low Dose Regime
    Zheng, Xiaoming
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2016, 6 (07) : 1818 - 1822
  • [33] Benchtop x-ray fluorescence computed tomography (XFCT) imaging
    Romero, Ignacio O.
    Fang, Yile
    Lun, Michael
    Li, Changqing
    MEDICAL IMAGING 2021: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2021, 11600
  • [34] Observation of parametric X-ray radiation by an imaging plate
    Takabayashi, Y.
    Shchagin, A. V.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2012, 278 : 78 - 81
  • [35] Radiation dose of digital radiography (DR) versus micro-dose x-ray (EOS) on patients with adolescent idiopathic scoliosis: 2016 SOSORT-IRSSD "John Sevastic Award" Winner in Imaging Research
    Hui, Steve C. N.
    Pialasse, Jean-Philippe
    Wong, Judy Y. H.
    Lam, Tsz-Ping
    Ng, Bobby K. W.
    Cheng, Jack C. Y.
    Chu, Winnie C. W.
    SCOLIOSIS AND SPINAL DISORDERS, 2016, 11
  • [36] Evaluation of CUDA for X-Ray Imaging System
    Lee, Seung Eun
    Park, Dae-Young
    COMPUTATIONAL INTELLIGENCE AND INTELLIGENT SYSTEMS, 2012, 316 : 621 - 625
  • [37] Hard x-ray imaging system for XEUS
    Kunieda, Hideyo
    Takahashi, Tadayuki
    Kokubun, Motohide
    Nakazawa, Kazuhiro
    Ogasaka, Yasushi
    SPACE TELESCOPES AND INSTRUMENTATION 2008: ULTRAVIOLET TO GAMMA RAY, PTS 1 AND 2, 2008, 7011
  • [38] Dose reduction in medical X-ray imaging using noise free photon counting
    Francke, T
    Eklund, M
    Ericsson, L
    Kristoffersson, T
    Peskov, V
    Rantanen, J
    Sokolov, S
    Söderman, J
    Ullberg, C
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 471 (1-2) : 85 - 87
  • [39] X-ray system simulation software tools for radiology and radiography education
    Kengyelics, Stephen M.
    Treadgold, Laura A.
    Davies, Andrew G.
    COMPUTERS IN BIOLOGY AND MEDICINE, 2018, 93 : 175 - 183
  • [40] Automated Inspection using X-ray Imaging
    Chen, Wenfei
    Miao, Zuohua
    Ming, Delie
    TRUSTCOM 2011: 2011 INTERNATIONAL JOINT CONFERENCE OF IEEE TRUSTCOM-11/IEEE ICESS-11/FCST-11, 2011, : 1769 - 1772