Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging system

被引:24
|
作者
Sharma, Reena [1 ]
Sharma, Sunil Dutt [1 ]
Pawar, Shramika [1 ]
Chaubey, Ajay [2 ]
Kantharia, S. [2 ]
Babu, D. A. R. [1 ]
机构
[1] CT & CRS, Radiol Phys & Advisory Div, Bombay, Maharashtra, India
[2] Bhabha Atom Res Ctr Hosp, Bombay, Maharashtra, India
关键词
Patient dose; entrance skin dose; diagnostic reference level; computed radiography; CR imaging system; X-ray machine; X-ray imaging;
D O I
10.4103/0971-6203.152244
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The screen-film system is replaced by computed radiography system for recording the images of the patients during X-ray radiography examinations. The change in imaging system requires the re-establishment of the institutional diagnostic reference levels (DRLs) for different types of X-ray examinations conducted at the hospital. For this purpose, patient specific parameters [age, height, weight, body mass index (BMI), object to image distance (OID)] and machine specific parameters (kVp, mAs, distance and field sizes) of 1875 patients during 21 different types of X-ray examinations were recorded for estimating the entrance skin dose (ESD). The ESD for each of these patients were estimated using measured X-ray beam output and the standard value of the back scatter factor. Five number summary was calculated for all the data for their presentation in the Box-Whisker plot, which provides the statistical distribution of the data. The data collected indicates that majorly performed examinations are cervical spine AP, Chest PA and Knee Lat with percentage contributions of 16.05, 16 and 8.27% respectively. The lowest contribution comes from Hip Lat which is about 1.01%. The ratio of measured ESD (maximum to minimum) for these examinations is found to be highest for the cervical spine AP with a value of 50 followed by Thoracic spine AP of 32.36. The ESD ratio for Chest PA, Knee Lat and Lumbar Spine AP are 30.75, 30.4 and 30.2 respectively. The lowest ESD ratio is for Hip Lat which is 2.68. The third quartile values of ESDs are established as the institutional DRLs. The ESD values obtained for 21 different X-ray projections are either comparable or lesser than the reported national/international values.
引用
收藏
页码:29 / 37
页数:9
相关论文
共 50 条
  • [21] X-ray imaging using the radiation diffracted by polycrystalline materials
    Wroblewski, T
    RADIATION PHYSICS AND CHEMISTRY, 2001, 61 (3-6) : 329 - 332
  • [22] Image acquisition of X-ray digital radiography system
    Cheng, YY
    Liu, X
    Wu, T
    Li, YH
    ISTM/2005: 6th International Symposium on Test and Measurement, Vols 1-9, Conference Proceedings, 2005, : 975 - 978
  • [23] X-ray imaging using digital cameras
    Winch, Nicola M.
    Edgar, Andrew
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [24] Fast X-Ray Luminescence Computed Tomography Imaging
    Liu, Xin
    Liao, Qimei
    Wang, Hongkai
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2014, 61 (06) : 1621 - 1627
  • [25] PATIENT DOSE AND IMAGE QUALITY IN 5 STANDARD X-RAY EXAMINATIONS
    HAVUKAINEN, R
    PIRINEN, M
    MEDICAL PHYSICS, 1993, 20 (03) : 813 - 817
  • [26] X-ray imaging and computed tomography for engineering applications
    Zabler, Simon
    Maisl, Michael
    Hornberger, Peter
    Hiller, Jochen
    Fella, Christian
    Hanke, Randolf
    TM-TECHNISCHES MESSEN, 2021, 88 (04) : 211 - 226
  • [27] Patient dose measurement in common medical X-ray examinations in Iran
    Rasuli, Behrouz
    Mahmoud-Pashazadeh, Ali
    Ghorbani, Mohammad
    Juybari, Raheleh Tabari
    Naserpour, Mozafar
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (01): : 374 - 386
  • [28] Implementation of a patient dose monitoring system in conventional digital X-ray imaging: initial experiences
    Heilmaier, Christina
    Zuber, Niklaus
    Weishaupt, Dominik
    EUROPEAN RADIOLOGY, 2017, 27 (03) : 1021 - 1031
  • [29] Fundamentals of X-ray imaging and radiation safety
    Bashore, T
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 54 (01) : 126 - 135
  • [30] Hybrid Monte Carlo and deterministic simulation approach for modeling a computed radiography imaging chain from X-ray exposure to optical readout
    Yao, Min
    Kaftandjian, Valerie
    Peterzol-Parmentier, Angela
    Schumm, Andreas
    Duvauchelle, Philippe
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2019, 941