Radiation dose to patients from X-ray radiographic examinations using computed radiography imaging system

被引:24
|
作者
Sharma, Reena [1 ]
Sharma, Sunil Dutt [1 ]
Pawar, Shramika [1 ]
Chaubey, Ajay [2 ]
Kantharia, S. [2 ]
Babu, D. A. R. [1 ]
机构
[1] CT & CRS, Radiol Phys & Advisory Div, Bombay, Maharashtra, India
[2] Bhabha Atom Res Ctr Hosp, Bombay, Maharashtra, India
关键词
Patient dose; entrance skin dose; diagnostic reference level; computed radiography; CR imaging system; X-ray machine; X-ray imaging;
D O I
10.4103/0971-6203.152244
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The screen-film system is replaced by computed radiography system for recording the images of the patients during X-ray radiography examinations. The change in imaging system requires the re-establishment of the institutional diagnostic reference levels (DRLs) for different types of X-ray examinations conducted at the hospital. For this purpose, patient specific parameters [age, height, weight, body mass index (BMI), object to image distance (OID)] and machine specific parameters (kVp, mAs, distance and field sizes) of 1875 patients during 21 different types of X-ray examinations were recorded for estimating the entrance skin dose (ESD). The ESD for each of these patients were estimated using measured X-ray beam output and the standard value of the back scatter factor. Five number summary was calculated for all the data for their presentation in the Box-Whisker plot, which provides the statistical distribution of the data. The data collected indicates that majorly performed examinations are cervical spine AP, Chest PA and Knee Lat with percentage contributions of 16.05, 16 and 8.27% respectively. The lowest contribution comes from Hip Lat which is about 1.01%. The ratio of measured ESD (maximum to minimum) for these examinations is found to be highest for the cervical spine AP with a value of 50 followed by Thoracic spine AP of 32.36. The ESD ratio for Chest PA, Knee Lat and Lumbar Spine AP are 30.75, 30.4 and 30.2 respectively. The lowest ESD ratio is for Hip Lat which is 2.68. The third quartile values of ESDs are established as the institutional DRLs. The ESD values obtained for 21 different X-ray projections are either comparable or lesser than the reported national/international values.
引用
收藏
页码:29 / 37
页数:9
相关论文
共 50 条
  • [1] Dose reference levels for examinations using computed radiography
    Gori, C
    Lazzari, B
    Rossi, F
    Zatelli, G
    Capaccioli, L
    Falivene, A
    Morandi, R
    Spanó, G
    PHYSICA MEDICA, 2001, 17 (02) : 93 - 96
  • [2] X-ray imaging plate using CsBr:Eu phosphors for computed radiography
    Nanto, H.
    Nishimura, A.
    Kuroda, M.
    Takei, Y.
    Nakano, Y.
    Shoji, T.
    Yanagita, T.
    Kasai, S.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01) : 278 - 281
  • [3] Effect of x-ray tube potential on image quality and patient dose for lumbar spine computed radiography examinations
    Brindhaban A.
    Al Khalifah K.
    Al Wathiqi G.
    Al Ostath H.
    Australasian Physics & Engineering Sciences in Medicine, 2005, 28 (4): : 216 - 222
  • [4] Comparison of image quality and patient dose for chest x-ray examinations on conventional and low cost computed radiography systems
    Muhogora, W. E.
    Trianni, A.
    Toso, F.
    Devetti, A.
    Padovani, R.
    Msaki, P.
    Kazema, R.
    RADIOGRAPHY, 2012, 18 (04) : 275 - 278
  • [5] Effects of x-ray spectra on the DQE of a computed radiography system
    Fetterly, KA
    Hangiandreou, NJ
    MEDICAL PHYSICS, 2001, 28 (02) : 241 - 249
  • [6] Effects of x-ray spectrum on the NPS of a computed radiography system
    Fetterly, KA
    Hangiandreou, NJ
    MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 : 632 - 639
  • [7] Radiation dose estimation for pencil beam X-ray luminescence computed tomography imaging
    Romero, Ignacio O.
    Li, Changqing
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2021, 29 (05) : 773 - 784
  • [8] Scale analysis using X-ray microfluorescence and computed radiography
    Candeias, J. P.
    de Oliveira, D. F.
    dos Anjos, M. J.
    Lopes, R. T.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 95 : 408 - 411
  • [9] Evaluation of digital flash X-ray images produced using computed radiography
    Carlsson, TE
    Örnhed, H
    26th International Congress on High Speed Photography and Photonics, 2005, 5580 : 842 - 850
  • [10] Determination of half value layers of X-ray equipment using computed radiography imaging plates
    Ariga, Eiji
    Ito, Shigeki
    Deji, Shizuhiko
    Saze, Takuya
    Nishizawa, Kunihide
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (01): : 71 - 75