KINETIC DETERMINATION OF COBALT AND NICKEL BY FLOW-INJECTION SPECTROPHOTOMETRY

被引:17
作者
ARRUDA, MAZ [1 ]
ZAGATTO, EAG [1 ]
MANIASSO, N [1 ]
机构
[1] UNIV SAO PAULO,CTR ENERGIA NUCL AGR,POB 96,BR-13400 PIRACICABA,SP,BRAZIL
基金
巴西圣保罗研究基金会;
关键词
FLOW INJECTION; KINETIC METHODS; UV VISIBLE SPECTROPHOTOMETRY; COBALT; NICKEL; STEELS;
D O I
10.1016/0003-2670(93)85259-M
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Reactor replacement is proposed as means of realizing different sample residence times in flow-injection analysis. As an application, an improved spectrophotometric determination of cobalt and nickel is proposed. The method exploits differences in reaction rates between cobalt and nickel citrate complexes and 4-(2-pyridylazo)resorcinol (PAR). The influence of pH, reagent concentrations, temperature, ionic strength, flow-rates and masking agent addition were investigated. Selectivity was enhanced by adding EDTA plus pyrophosphate and exploiting kinetic discrimination. The system is very stable and allows the determination of cobalt and nickel up to 5.00 mg l-1 with a sampling frequency of 40 h-1. For tool steel samples characterized by low nickel/cobalt concentration ratios, the system can be simplified, allowing kinetic masking of nickel and single cobalt determination. The reagent consumption is 65.3 mug of PAR per determination, and precise results (R.S.D. < 1%) in agreement with inductively coupled plasma atomic emission spectrometry are obtained at a sampling frequency of 100 h-1.
引用
收藏
页码:476 / 480
页数:5
相关论文
共 13 条
[1]  
[Anonymous], 1987, ANALYST, V112, P199
[2]   Catalytic Determination of Cobalt in Liver by Flow-Injection Spectrophotometry [J].
Arruda, M. A. Z. ;
Zagatto, E. A. G. ;
Jacinth, A. O. ;
Brienza, S. M. B. .
JOURNAL OF THE BRAZILIAN CHEMICAL SOCIETY, 1991, 2 (01) :47-50
[3]   A SIMPLE STOPPED-FLOW METHOD WITH CONTINUOUS PUMPING FOR THE SPECTROPHOTOMETRIC FLOW-INJECTION DETERMINATION OF BORON IN PLANTS [J].
ARRUDA, MAZ ;
ZAGATTO, EAG .
ANALYTICA CHIMICA ACTA, 1987, 199 :137-145
[4]   2 POINT KINETIC SIMULTANEOUS DETERMINATION OF COBALT(II) AND NICKEL(II) IN AQUEOUS-SOLUTION USING FLOW-INJECTION ANALYSIS (FIA) [J].
BETTERIDGE, D ;
FIELDS, B .
FRESENIUS ZEITSCHRIFT FUR ANALYTISCHE CHEMIE, 1983, 314 (04) :386-390
[5]   SIMULTANEOUS DETERMINATIONS IN FLOW-INJECTION ANALYSIS - A REVIEW [J].
DECASTRO, MDL ;
CASES, MV .
ANALYST, 1984, 109 (04) :413-419
[6]  
DECASTRO MDL, 1986, TRAC-TREND ANAL CHEM, V5, P71
[7]   COMPARISON OF FLOW-INJECTION ANALYSIS CONFIGURATIONS FOR DIFFERENTIAL KINETIC DETERMINATION OF COBALT AND NICKEL [J].
FERNANDEZ, A ;
DECASTRO, MDL ;
VALCARCEL, M .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :1146-1151
[8]  
MOTTOLA HA, 1988, ANAL CHEM, V60, pR181, DOI 10.1021/ac00163a014
[9]  
MOTTOLA HA, 1984, ANAL CHEM, V56, pR96
[10]  
Ringbom A., 1963, CHEM ANAL, VXVI