ELECTROSPRAY IONIZATION MASS-SPECTRA OF HEMOGLOBIN AND TRANSFERRIN BY A MAGNETIC-SECTOR MASS-SPECTROMETER - COMPARISON WITH THEORETICAL ISOTOPIC DISTRIBUTIONS

被引:31
作者
WADA, Y
TAMURA, J
MUSSELMAN, BD
KASSEL, DB
SAKURAI, T
MATSUO, T
机构
[1] JEOL USA INC,PEABODY,MA 01960
[2] MASSACHUSETTS GEN HOSP,DEPT MOLEC BIOL,BOSTON,MA 01960
[3] OSAKA UNIV,COLL GEN EDUC,INST PHYS,TOYONAKA,OSAKA 560,JAPAN
关键词
D O I
10.1002/rcm.1290060104
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Electrospray ionization mass spectra of human hemoglobin chains and of transferrin were acquired on a magnetic sector mass spectrometer. The observed molecular ion for each hemoglobin chain was in good agreement with the theoretical isotopic distribution at a reasonable resolution of 2000. The clear separation of a variant beta-chain in admixture with the normal counterpart at mass 15 867 that differed from it by 14 Da (0.09%) ensured that a smaller mass difference could be detected. The molecular ions for human transferrin were too broad compared with the theoretical shape to determine the molecular mass accurately, probably due to the heterogeneity of the carbohydrate moiety. A decrease in mass by neuraminidase digestion, however, determined the average number of sialic acids in the molecule.