ANALYSIS OF THE SPATIAL-FREQUENCY-DEPENDENT DQE OF OPTICALLY COUPLED DIGITAL MAMMOGRAPHY DETECTORS

被引:65
|
作者
MAIDMENT, ADA
YAFFE, MJ
机构
[1] UNIV TORONTO,DEPT MED BIOPHYS,TORONTO M4N 3M5,ON,CANADA
[2] SUNNYBROOK HLTH SCI CTR,DIV RES,TORONTO M4N 3M5,ON,CANADA
[3] UNIV TORONTO,DEPT RADIOL,TORONTO M4N 3M5,ON,CANADA
关键词
DIGITAL MAMMOGRAPHY; X-RAY DETECTOR; OPTICAL COUPLING EFFICIENCY; MODULATION TRANSFER FUNCTION; NOISE POWER SPECTRA; DETECTIVE QUANTUM EFFICIENCY;
D O I
10.1118/1.597331
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of optical coupling efficiency on the spatial-frequency-dependent propagation of signal and noise is considered for x-ray image detectors for digital mammography in which a phosphor screen is optically coupled to a charge-coupled device (CCD) image array. For experimental purposes, optical coupling between a Gd2O2S:Tb phosphor screen and a CCD image array was provided by relay lenses. Neutral density filters were inserted between the lenses to vary the optical coupling efficiency without altering the inherent spatial resolution. The total coupling efficiency, defined as the number of electrons (e-) recorded in the CCD per x-ray interaction in the phosphor, was calculated in each case. The modulation transfer function, and the contributions to the total noise power spectrum (NPS) of x-ray quantum noise, secondary quantum noise, and inherent detector noise were measured as a function of coupling efficiency. These data were used to calculate the spatial-frequency-dependent detective quantum efficiency [DOE(f)]. The NPS due to x-ray quantum noise had a significant spatial-frequency dependence for coupling efficiencies of more than 9 e- per x-ray interaction, but little spatial-frequency dependence for coupling efficiencies of less than 2 e- per x-ray interaction. These results indicate that to preserve high spatial-frequency values of DQE(f), and to ensure that images are x-ray quantum-noise limited at high spatial frequencies, a coupling efficiency on the order of 10 e- per x-ray interaction is required. This is contrary to the common belief, based on zero spatial-frequency analysis, that a coupling efficiency on the order of 1 e- per x-ray interaction is sufficient to acquire x-ray quantum-noise-limited images.
引用
收藏
页码:721 / 729
页数:9
相关论文
共 50 条
  • [1] An experimental method for the determination of spatial-frequency-dependent detective quantum efficiency (DQE) of scintillators used in X-ray imaging detectors
    Kandarakis, I
    Cavouras, D
    Panayiotakis, GS
    Triantis, D
    Nomicos, CD
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 399 (2-3): : 335 - 342
  • [2] Spatial frequency dependent DQE performance of a CsI:Tl based X-ray detector for digital mammography
    Jing, ZX
    Huda, W
    Walker, JK
    Choi, WY
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 159 - 168
  • [3] DQE of imaging detectors in terms of spatial frequency
    Zanella, G
    Zannoni, R
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 437 (01): : 163 - 167
  • [4] ANALYSIS OF SIGNAL PROPAGATION IN OPTICALLY COUPLED DETECTORS FOR DIGITAL MAMMOGRAPHY .1. PHOSPHOR SCREENS
    MAIDMENT, ADA
    YAFFE, MJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05): : 877 - 889
  • [5] Analysis of factors that affect DQE in digital mammography
    Takamura, M
    Higaki, A
    Kodera, Y
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 1217 - 1228
  • [6] Analysis of signal propagation in optically coupled detectors for digital mammography .2. Lens and fibre optics
    Maidment, ADA
    Yaffe, MJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (03): : 475 - 493
  • [7] Validation of a Custom-made Software for DQE Assessment in Mammography Digital Detectors
    Ayala-Dominguez, L.
    Perez-Ponce, H.
    Brandan, M. E.
    ELEVENTH MEXICAN SYMPOSIUM ON MEDICAL PHYSICS, 2010, 1310 : 39 - 43
  • [8] SPATIAL-FREQUENCY-DEPENDENT BRIGHTNESS INDUCTION EFFECT
    MCCOURT, ME
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1980, 70 (12) : 1599 - 1599
  • [9] Determination of DQE as a quantitative assessment of detectors in digital mammography: Measurements and calculation in practice
    Wysocka-Rabin, Anna
    Dobrzynska, Magdalena
    Pasicz, Katarzyna
    Skrzynski, Witold
    Fabiszewska, Ewa
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2021, 27 (03): : 223 - 232
  • [10] Application of DQE for quantitative assessment of detectors to estimate AEC efficiency in digital mammography
    Fabiszewska, Ewa
    Wysocka-Rabin, Anna
    Dobrzynska, Magdalena
    Skrzynski, Witold
    Pasicz, Katarzyna
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2021, 27 (01): : 51 - 56