ALIGNMENT OF TOTAL KNEE ARTHROPLASTY - THE RELATIONSHIP TO RADIOLUCENCY AROUND THE TIBIAL COMPONENT

被引:14
作者
HARVEY, IA
MANNING, MP
SAMPATH, SAC
JOHNSON, R
ELLOY, MA
机构
[1] BUPA, MURRAYFIELD HOSP, WIRRAL KNEE CTR, KNEE RES UNIT, WIRRAL L61 1AU, MERSEYSIDE, ENGLAND
[2] WALTON HOSP, LIVERPOOL 9, MERSEYSIDE, ENGLAND
[3] WHISTON HOSP, N PRESCOT, LANCS, ENGLAND
[4] UNIV LIVERPOOL, DEPT ORTHOPAED SURG, LIVERPOOL L69 3BX, MERSEYSIDE, ENGLAND
[5] UNIV LIVERPOOL, DEPT MECH ENGN, LIVERPOOL L69 3BX, MERSEYSIDE, ENGLAND
关键词
TOTAL KNEE REPLACEMENT; ALIGNMENT; TIBIAL LOOSENING;
D O I
10.1016/1350-4533(95)95708-I
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Between January 1982 and July 1985, 122 cemented 'Accord' total Knee replacements were performed Out of this number, 21 were lost to follow-up because of death (in 16), revision (2) due to infection, and failure to attend (3). The post-operative alignment of 101 prostheses were compared to the incidence of radiolucencies around the tibial component at 5 Bears follow-up. The mechanical axis was used as a reference. The mean alignment was 0.99 degrees valgus with a standard deviation of 2.48 degrees. Some 72% of knees were within 3 degrees and 94% within 5 degrees of true alignment. Using two methods of assessing radiolucencies there was a non-significant relationship, between the alignment and radiolucencies. The alignment tolerance with tills prosthesis is, therefore, at least 5 degrees.
引用
收藏
页码:182 / 187
页数:6
相关论文
共 12 条
[1]  
BARGREN JH, 1983, CLIN ORTHOP RELAT R, P178
[2]  
CAMERON HU, 1982, CLIN ORTHOP RELAT R, V170, P141
[3]  
ECKER ML, 1987, CLIN ORTHOP RELAT R, V216, P151
[4]   ACCURACY OF INTRAMEDULLARY ALIGNMENT IN TOTAL KNEE REPLACEMENT [J].
ELLOY, MA ;
MANNING, MP ;
JOHNSON, R .
JOURNAL OF BIOMEDICAL ENGINEERING, 1992, 14 (05) :363-370
[5]  
EWALD FC, 1989, CLIN ORTHOP RELAT R, P9
[6]  
INSALL JN, 1984, SURGERY KNEE
[7]   CORONAL ALIGNMENT AFTER TOTAL KNEE REPLACEMENT [J].
JEFFERY, RS ;
MORRIS, RW ;
DENHAM, RA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (05) :709-714
[8]   POLYCENTRIC TOTAL KNEE ARTHROPLASTY - A 10-YEAR FOLLOW-UP-STUDY [J].
LEWALLEN, DG ;
BRYAN, RS ;
PETERSON, LFA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (08) :1211-1218
[9]  
Morrison J B, 1969, Biomed Eng, V4, P573
[10]  
SMITH JL, 1989, J ARTHR S55, V4