Needle placement errors: do we need steerable needles in interventional radiology?

被引:38
作者
de Jong, Tonke L. [1 ]
van de Berg, Nick J. [1 ]
Tas, Lisette [1 ]
Moelker, Adriaan [2 ]
Dankelman, Jenny [1 ]
van den Dobbelsteen, John J. [1 ]
机构
[1] Delft Univ Technol, BioMech Engn Dept, Mekelweg 2, NL-2628 CD Delft, Netherlands
[2] Univ Med Ctr, Radiol & Nucl Med Dept, Erasmus MC, Rotterdam, Netherlands
来源
MEDICAL DEVICES-EVIDENCE AND RESEARCH | 2018年 / 11卷
关键词
clinical use; interventional radiology; needle bending; needle deflection; needle placement error; questionnaire; steerable needle;
D O I
10.2147/MDER.S160444
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: Accurate and precise needle placement is of utmost importance in interventional radiology. However, targeting can be challenging due to, eg, tissue motion and deformation. Steerable needles are a possible solution to overcome these challenges. The present work studied the clinical need for steerable needles. We aimed to answer three subquestions: 1) What are the current challenges in needle placement? 2) What are allowable needle placement errors? and 3) Do current needles need improvement and would steerable needles add clinical value? Methods: A questionnaire was administered at the Annual Meeting of Cardiovascular and Interventional Radiology Society of Europe in 2016. In total, 153 respondents volunteered to fill out the survey, among them 125 (interventional) radiologists with experience in needle placement. Results: 1) Current challenges in needle placement include patient-specific and technical factors. Movement of the target due to breathing makes it most difficult to place a needle (90%). 2) The mean maximal allowable needle placement error in targeted lesions is 2.7 mm. A majority of the respondents (85%) encounter unwanted needle bending upon insertion. The mean maximal encountered unwanted needle bending is 5.3 mm. 3) Needles in interventional radiology need improvement, eg, improved needle visibility and manipulability, according to 95% of the respondents. Added value for steerable needles in current interventions is seen by 93% of the respondents. Conclusion: Steerable needles have the potential to add clinical value to radiologic interventions. The current data can be used as input for defining clinical design requirements for technical tools, such as steerable needles and navigation models, with the aim to improve needle placement in interventional radiology.
引用
收藏
页码:259 / 265
页数:7
相关论文
共 18 条
[1]  
Cowan NJ, 2011, SURGICAL ROBOTICS: SYSTEMS APPLICATIONS AND VISIONS, P557, DOI 10.1007/978-1-4419-1126-1_23
[2]   Concentric Tube Robots as Steerable Needles: Achieving Follow-the-Leader Deployment [J].
Gilbert, Hunter B. ;
Neimat, Joseph ;
Webster, Robert J., III .
IEEE TRANSACTIONS ON ROBOTICS, 2015, 31 (02) :246-258
[3]   Image Guidance of Flexible Tip-Steerable Needles [J].
Kallem, Vinutha ;
Cowan, Noah J. .
IEEE TRANSACTIONS ON ROBOTICS, 2009, 25 (01) :191-196
[4]  
Kratchman LB, 2011, P SOC PHOTOGRAPHIC I
[5]   Behavior of Tip-Steerable Needles in Ex Vivo and In Vivo Tissue [J].
Majewicz, Ann ;
Marra, Steven P. ;
van Vledder, Mark G. ;
Lin, MingDe ;
Choti, Michael A. ;
Song, Danny Y. ;
Okamura, Allison M. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2012, 59 (10) :2705-2715
[6]   Driven Around the Bend: Novel Use of a Curved Steerable Needle [J].
Murphy, Darra T. ;
Korzan, Jeffrey R. ;
Ouellette, Hugue A. ;
Liu, David M. ;
Clarkson, Paul W. ;
Munk, Peter L. .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 36 (02) :531-535
[7]   Ultrasound-Guided Robot for Flexible Needle Steering [J].
Neubach, Zipi ;
Shoham, Moshe .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (04) :799-805
[8]   Hand-held steerable needle device [J].
Okazawa, S ;
Ebrahimi, R ;
Chuang, J ;
Salcudean, SE ;
Rohling, R .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2005, 10 (03) :285-296
[9]   A novel curvilinear approach for prostate seed implantation [J].
Podder, Tarun K. ;
Dicker, Adam P. ;
Hutapea, Parsaoran ;
Darvish, Kurosh ;
Yu, Yan .
MEDICAL PHYSICS, 2012, 39 (04) :1887-1892
[10]   Robot-Assisted Needle Steering [J].
Reed, Kyle B. ;
Majewicz, Ann ;
Kallem, Vinutha ;
Alterovitz, Ron ;
Goldberg, Ken ;
Cowan, Noah J. ;
Okamura, Allison M. .
IEEE ROBOTICS & AUTOMATION MAGAZINE, 2011, 18 (04) :35-46