FLUORESCENCE-LABELED PEPTIDES AS ISOELECTRIC POINT (PI) MARKERS IN CAPILLARY ISOELECTRIC-FOCUSING WITH FLUORESCENCE DETECTION

被引:24
作者
SHIMURA, K
KASAI, K
机构
[1] Faculty of Pharmaceutical Sciences, Teikyo University, Sagamiko, Kanagawa
关键词
FLUORESCENT PI MARKER; ISOELECTRIC FOCUSING; CAPILLARY ELECTROPHORESIS; LASER-INDUCED FLUORESCENCE; FLUORESCENCE-LABELED PEPTIDES;
D O I
10.1002/elps.11501601245
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Commercially available peptides, mostly angiotensin derivatives, were labeled at their N-terminal amino group with 5-carboxytetramethylrhodamine succinimidyl ester, to obtain fluorescent pI markers for capillary isoelectric focusing with fluorescence detection. The labeled peptides were purified by reversed-phase chromatography. They were well separated on isoelectric focusing in a polyacrylamide gel slab and their pIs were determined by comigration with protein-pI markers. The fluorescent markers could be detected as sharp peaks in capillary isoelectric focusing with laser-induced fluorescence detection (He-Ne laser, 1 mW 543.5 nm). The detection limit was found to be around 3 X 10(-12) M (0.8 amol). Tetramethylrhodamine-labeled pea lectin (3 pg) was subjected to capillary isoelectric focusing and the pIs of the fluorescent derivatives of the lectin were determined by using the fluorescence-labeled peptides as pI markers.
引用
收藏
页码:1479 / 1484
页数:6
相关论文
共 15 条