A HIGH-RESOLUTION SPECT SYSTEM BASED ON A MICROCHANNEL-PLATE IMAGER

被引:15
作者
DILMANIAN, FA
WEBER, DA
CODERRE, JA
JOEL, DD
SHI, KC
MEINKEN, GE
SOM, P
TANG, YN
VOLKOW, ND
YEE, C
BRILL, AB
WATANABE, M
INUZUKA, E
OBA, K
GERSON, R
IIDA, H
HIRUMA, A
机构
[1] HAMAMATSU PHOTON KK,HAMAMATSU,JAPAN
[2] HAMAMATSU PHOTON SYST,BRIDGEWATER,NJ
关键词
D O I
10.1109/23.106699
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We have developed a high-resolution single-photon emission-computed tomography (SPECT) system for small-animal studies with 125I. The SPECT camera uses a Hamamatsu microchannel-plate imager PIAS (Photon-Counting Image Acquisition System), a 0.7 mm thick NaI(Tl) plate, and a lead-glass parallel-hole collimator. The 15-mm diameter field-of-view (FOV) of the PIAS tube was enlarged by an optical fiber face-plate taper to provide a 26 mm camera FOV. Camera absolute sensitivity (counts detected/gamma photons emitted) is 2×10−5 for 125I. The system provides an in-plane resolution of about 0.7 mm FWHM in the reconstructed image for a 15-mm radius of rotation. SPECT images from phantom studies and from organs in the rat and mouse provide examples of the resolving capacity of the system. Copyright © 1990 by The Institute of Electrical and Electronics Engineers, Inc.
引用
收藏
页码:687 / 695
页数:9
相关论文
共 7 条
  • [1] BIZAIS YJC, 1983, UNPUB HOLPLT
  • [2] DILMANIAN FA, 1988, INVEST RADIOL, V22, pS7
  • [3] HUESMAN, 1977, UNPUB RECLBL LIBRARY
  • [4] KINOSHITA M, 1985, ADV IMAG ELECTR PHYS, V64, P323
  • [5] KINOSHITA M, 1985, PIAS CR3000 PHOT COU
  • [6] TSUCHIYA Y, 1985, ADV ELECTRON EL PHYS, V64, P21
  • [7] TUSCHIYA Y, 1985, J IMAGE TECH, V11, P215