3-DIMENSIONAL OPTICAL-TRANSFER-FUNCTION ANALYSIS FOR A LASER-SCAN FLUORESCENCE MICROSCOPE WITH AN EXTENDED DETECTOR

被引:38
作者
KAWATA, S
ARIMOTO, R
NAKAMURA, O
机构
[1] Department of Applied Physics, Osaka University Suita, Osaka
[2] Instruments Division, Nikon Corporation, Yokohama
[3] National Research Laboratory of Metrology, Tsukuba
来源
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION | 1991年 / 8卷 / 01期
关键词
D O I
10.1364/JOSAA.8.000171
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Three-dimensional imaging characteristics of a laser-scan fluorescence microscope with various sizes of detector are quantitatively analyzed through the studies and calculations of the three-dimensional optical transfer function of the microscope. It is found that a detector of an Airy-disk size formed by the objective lens in the detector plane provides both reasonable energy-collection efficiency and three-dimensionally resolved imaging capability. It is also found that a slit detector provides longitudinal resolution of the sample structure similarly to a confocal microscope, while the lateral resolution is approximately the same as that of a conventional microscope.
引用
收藏
页码:171 / 175
页数:5
相关论文
共 15 条