A SURVEY OF TECHNOLOGY AND TEMPERATURE CONTROL ON A NEONATAL UNIT IN KATHMANDU, NEPAL

被引:18
作者
JOHANSON, RB
MALLA, DS
TULADHAR, C
AMATYA, M
SPENCER, SA
ROLFE, P
机构
[1] KATHMANDU MATERN HOSP,KATMANDU,NEPAL
[2] N STAFFORDSHIRE MATERN HOSP,DEPT NEONATOL,STOKE ON TRENT ST4 6SD,ENGLAND
[3] TRIBHUVAN UNIV,DEPT BIOMED ENGN,KATMANDU,NEPAL
[4] UNIV KEELE,DEPT BIOMED ENGN,KEELE ST5 5BG,STAFFS,ENGLAND
[5] N STAFFORDSHIRE MATERN HOSP,DEPT OBSTET,STOKE ON TRENT ST4 6SD,ENGLAND
关键词
D O I
10.1093/tropej/39.1.4
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
In assessment of the incubators in use at the Kathmandu Maternity Hospital neonatal unit was undertaken; this Aas followed by a prospective survey of neonatal temperatures on the unit. In the incubator assessment 11 studies were carried out in five incubators. Three of the thermostats in the five incubators did not work at all and those in the other two incubators were more than 3-degrees-C inaccurate. All the incubator thermometers gave recordings less than the actual temperature (with a range of error: 1.3-4.4-degrees-C). Six out of the 11 babies studied were hypothermic (defined as rectal temperature < 36-degrees-C). In the prospective surVey of temperatures a high incidence of hypothermia was found on the neonatal unit at the time of first temperature measurement (64 per cent had a rectal temperature of < 36-degrees-C). A significant association between admission hypothermia and mortality was noted. Sixteen per cent (10/64) of babies admitted with temperatures < 36-degrees-C died within the first week, compared to 0 (0/36) of those admitted with temperatures greater-than-or-equal-to 36-degrees-C.
引用
收藏
页码:4 / 10
页数:7
相关论文
共 27 条
  • [1] OXYGENATION, HEART-RATE AND TEMPERATURE IN VERY LOW-BIRTH-WEIGHT INFANTS DURING SKIN-TO-SKIN CONTACT WITH THEIR MOTHERS
    ACOLET, D
    SLEATH, K
    WHITELAW, A
    [J]. ACTA PAEDIATRICA SCANDINAVICA, 1989, 78 (02): : 189 - 193
  • [2] BARNES SE, 1977, LANCET, V1, P544
  • [3] BOLLIGER P, 1977, APPROPRIATE TECH, V4, P14
  • [4] BUETOW KC, 1964, PEDIATRICS, V34, P163
  • [5] BACTERIAL COLONIZATION OF INFANTS RAISED IN INCUBATORS AND UNDER RADIANT HEATERS
    CHANG, CT
    GLASS, L
    EVANS, HE
    PIEROG, SH
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1977, 52 (06) : 507 - 509
  • [6] CHINTU C, 1978, MED J ZAMBIA, V12, P110
  • [7] APPROPRIATE-TECHNOLOGY IN KEEPING BABIES WARM IN INDIA
    DAGA, SR
    CHANDRASHEKHAR, L
    POL, PP
    PATOLE, S
    [J]. ANNALS OF TROPICAL PAEDIATRICS, 1986, 6 (01): : 23 - 25
  • [8] DAY RL, 1964, PEDIATRICS, V34, P171
  • [9] DELEEUW R, 1988, NED TIJDSCHR GENEESK, V131, P1484
  • [10] DOUEK E, 1976, LANCET, V1, P1110