A MODEL FOR ESTIMATING CORNEAL STIFFNESS USING AN INDENTER

被引:14
作者
CARNELL, PH
VITO, RP
机构
[1] School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, GA
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 1992年 / 114卷 / 04期
关键词
D O I
10.1115/1.2894111
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A finite element model that simulates indentation of the cornea was generated to examine the feasibility of using indentation data to determine mechanical properties. A layered model which included geometric nonlinearities was necessary to characterize the indentation process. Results indicate that a secant modulus could be determined by measuring indenter force and contact area.
引用
收藏
页码:549 / 552
页数:4
相关论文
共 23 条
  • [1] BRYANT M R, 1987, CLAO Journal, V13, P238
  • [2] BRYANT MR, 1989, ADV DESIGN AUTOMATIO, P363
  • [3] CARNELL PH, 1990, FINITE ELEMENT MODEL
  • [4] Cavanagh, 1988, CORNEA, P187
  • [5] COLLINS R, 1980, MATH MODELS DYNAMICS
  • [6] Friedenwald JS., 1937, AM J OPHTHALMOL, V20, P985, DOI [10.1016/S0002-9394(37)90425-2, DOI 10.1016/S0002-9394(37)90425-2]
  • [7] CORNEAL CURVATURE CHANGE DUE TO STRUCTURAL ALTERNATION BY RADIAL KERATOTOMY
    HUANG, T
    BISARNSIN, T
    SCHACHAR, RA
    BLACK, TD
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1988, 110 (03): : 249 - 253
  • [8] THE MECHANICAL-PROPERTIES OF THE RABBIT AND HUMAN CORNEA
    JUE, B
    MAURICE, DM
    [J]. JOURNAL OF BIOMECHANICS, 1986, 19 (10) : 847 - 853
  • [9] KALNINS A, 1979, APPL PHYSL MECH, P731
  • [10] ANALYSIS OF CORNEO-SCLERAL SHELL BY METHOD OF DIRECT STIFFNESS
    KOBAYASHI, AS
    WOO, SLY
    LAWRENCE, C
    SCHLEGEL, WA
    [J]. JOURNAL OF BIOMECHANICS, 1971, 4 (05) : 323 - +