A new analytical formula for neutron capture gamma dose calculations in double-bend mazes in radiation therapy

被引:11
作者
Ghiasi, Hosein [1 ]
Mesbahi, Asghar [1 ,2 ]
机构
[1] Tabriz Univ Med Sci, Med Sch, Med Phys Dept, Tabriz, Iran
[2] Imam Hosp, Radiat Oncol Dept, Tabriz, Iran
关键词
Capture gamma ray; Double-bend maze; Radiation therapy bunkers; Radiotherapy shielding;
D O I
10.1016/j.rpor.2012.03.011
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Photoneutrons are produced in radiation therapy with high energy photons. Also, capture gamma rays are the byproduct of neutrons interactions with wall material of radiotherapy rooms. Aim: In the current study an analytical formula was proposed for capture gamma dose calculations in double bend mazes in radiation therapy rooms. Materials and methods: A total of 40 different layouts with double-bend mazes and a 18 MeV photon beam of Varian 2100 Clinac were simulated using MCNPX Monte Carlo (MC) code. Neutron capture gamma ray dose equivalent was calculated by the MC method along the maze and at the maze entrance door of all the simulated rooms. Then, all MC resulted data were fitted to an empirical formula for capture gamma dose calculations. Wu-McGinley analytical formula for capture gamma dose equivalent at the maze entrance door in single-bend mazes was also used for comparison purposes. Results: For capture gamma dose equivalents at the maze entrance door, the difference of 2-11% was seen between MC and the derived equation, while the difference of 36-87% was found between MC and the Wu-McGinley methods. Conclusion: Our results showed that the derived formula results were consistent with the MC results for all of 40 different geometries. However, as a new formula, further evaluations are required to validate its use in practical situations. Finally, its application is recommend for capture gamma dose calculations in double-bend mazes to improve shielding calculations. (C) 2012 Greater Poland Cancer Centre. Published by Elsevier Urban & Partner Sp. z.o.o. All rights reserved.
引用
收藏
页码:220 / 225
页数:6
相关论文
共 21 条
[1]   Photoneutron intensity variation with field size around radiotherapy linear accelerator 18-MeV X-ray beam [J].
Al-Ghamdi, H. ;
Fazal-ur-Rehman ;
Al-Jarallah, M. I. ;
Maalej, N. .
RADIATION MEASUREMENTS, 2008, 43 (SUPPL.1) :S495-S499
[2]   Thermoluminescence measurements of neutron dose around a medical linac [J].
Barquero, R ;
Méndez, R ;
Iñiguez, MP ;
Vega, HR ;
Voytchev, M .
RADIATION PROTECTION DOSIMETRY, 2002, 101 (1-4) :493-496
[3]   Photonuclear dose calculations for high-energy photon beams from Siemens and Varian linacs [J].
Chibani, O ;
Ma, CMC .
MEDICAL PHYSICS, 2003, 30 (08) :1990-2000
[4]  
Ghiasi H, 2010, IRAN J RADIAT RES, V8, P187
[5]   Measurements of secondary neutron dose from 15 MV and 18 MV IMRT [J].
Howell, RM ;
Ferenci, MS ;
Hertel, NE ;
Fullerton, GD ;
Fox, T ;
Davis, LW .
RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) :508-512
[6]  
IAEA, 2006, IAEA SAF REP SER, V47
[7]   Out-of-field photon and neutron dose equivalents from step-and-shoot intensity-modulated radiation therapy [J].
Kry, SF ;
Salehpour, M ;
Followill, DS ;
Stovall, M ;
Kuban, DA ;
White, RA ;
Rosen, II .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1204-1216
[8]  
LARIVIERE PD, 1986, RADIAT PROT DOSIM, V14, P257
[9]   EVALUATION OF NEUTRON DOSE EQUIVALENT LEVELS AT THE MAZE ENTRANCE OF MEDICAL ACCELERATOR TREATMENT ROOMS [J].
MCGINLEY, PH ;
BUTKER, EK .
MEDICAL PHYSICS, 1991, 18 (02) :279-281
[10]   Photoneutron and capture gamma dose equivalent for different room and maze layouts in radiation therapy [J].
Mesbahi, Asghar ;
Ghiasi, Hosein ;
Mahdavi, Seyed Rabee .
RADIATION PROTECTION DOSIMETRY, 2010, 140 (03) :242-249