Evaluation of the performance of parallel-hole collimator for high resolution small animal SPECT: A Monte Carlo study

被引:0
|
作者
Sadremomtaz, Alireza [1 ]
Telikani, Zeinab [1 ]
机构
[1] Univ Guilan, Dept Phys, Fac Sci, Rasht, Iran
来源
IRANIAN JOURNAL OF NUCLEAR MEDICINE | 2016年 / 24卷 / 02期
关键词
SPECT; Compact pixelated gamma camera; Septal penetration; Scatter; Monte Carlo simulation;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Image quality and accuracy of in vivo activity quantification in SPECT are affected by collimator penetration and scatter components, especially in high energy imaging. These phenomena highly depend on the collimator characteristic and photon energy. The presence of penetrated and scattered photons from collimator in SPECT images degrades spatial resolution, contrast and image quality. Knowledge of penetration and scatter distribution is essential for optimization of collimator design and development of reconstruction algorithms. The aim of this study to survey the collimator performance of the newly developed HiReSPECT dual head gamma camera with pixelated array CsI(Na). Methods: We modeled the HiReSPECT, by using SIMIND Monte Carlo simulation code. The contribution of geometric, scatter and penetration components were quantitatively calculated for the different energy sources. Then we compared these results with simulation results of another small animal SPECT with compact pixelated array CsI(Tl) detector. Results: The simulated System spatial resolution and energy resolution of the HiReSPECT at 140keV respectively are 1.9mm and 29.72 keV (21.23%) FWHM at 2.5cm distance from detector surface also Geometric, penetration, and scatter at 140keV for the HiReSPECT collimator are 96.42%, 2.22%, 1.30%, respectively. Similarly, geometric, penetration, and scatter at 159keV and 245keV for this system collimator are (95.24%, 3.08%, 1.68%) and (87.21%, 8.10%, 4.69%), respectively. Conclusion: The results verified that the magnitude of these components depend on collimator geometric structure and photons energy. The measured performances indicated that the HiReSPECT scanner is well suited for preclinical molecular imaging research and provide high resolution for small animal imaging.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 50 条
  • [1] A Monte Carlo study on the performance evaluation of a parallel hole collimator for a HiReSPECT: A dedicated small-animal SPECT
    Abbaspour, Samira
    Tanha, Kaveh
    Mahmoudian, Babak
    Assadi, Majid
    Islamian, Jalil Pirayesh
    APPLIED RADIATION AND ISOTOPES, 2018, 139 : 53 - 60
  • [2] A Monte Carlo simulation study of the feasibility of a high resolution parallel-hole collimator with a CdTe pixelated semiconductor SPECT system
    Lee, Y. -J.
    Park, S. -J.
    Lee, S. -W.
    Kim, D. -H.
    Kim, Y. -S.
    Jo, B. -D.
    Kim, H. -J.
    JOURNAL OF INSTRUMENTATION, 2013, 8
  • [3] Monte carlo simulation study of a novel parallel-hole collimator with a CdTe pixelated semiconductor SPECT system
    Lee, Young-Jin
    Kim, Dae-Hong
    Rhee, Yong-Chun
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (11) : 1737 - 1744
  • [4] Monte carlo simulation study of a novel parallel-hole collimator with a CdTe pixelated semiconductor SPECT system
    Young-Jin Lee
    Dae-Hong Kim
    Yong-Chun Rhee
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2014, 64 : 1737 - 1744
  • [5] Characterization of parallel-hole collimator using Monte Carlo Simulation
    Pandey, Anil Kumar
    Sharma, Sanjay Kumar
    Karunanithi, Sellam
    Kumar, Praveen
    Bal, Chandrasekhar
    Kumar, Rakesh
    INDIAN JOURNAL OF NUCLEAR MEDICINE, 2015, 30 (02): : 128 - 134
  • [6] The Effect of High-resolution Parallel-hole Collimator Materials with a Pixelated Semiconductor SPECT System at Equivalent Sensitivities: Monte Carlo Simulation Studies
    Lee, Young-Jin
    Kim, Dae-Hong
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 64 (07) : 1055 - 1062
  • [7] The effect of high-resolution parallel-hole collimator materials with a pixelated semiconductor SPECT system at equivalent sensitivities: Monte Carlo simulation studies
    Young-Jin Lee
    Dae-Hong Kim
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2014, 64 : 1055 - 1062
  • [8] Performance evaluation of high-resolution square parallel-hole collimators with a CZT room temperature pixelated semiconductor SPECT system: a Monte Carlo simulation study
    Lee, Y.
    Kang, W.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [9] Optimized Parallel-Hole Collimator Design that Balances Spatial Resolution and Sensitivity through Monte Carlo Simulation
    Kang, Seung-Hun
    Jo, Byungdu
    Lee, Seung-Jae
    JOURNAL OF MAGNETICS, 2023, 28 (04) : 415 - 419
  • [10] Performance evaluation of a parallel-hole collimated detector module for animal SPECT imaging
    黄先超
    章志明
    李道武
    唐浩辉
    李婷
    廖燕飞
    刘军辉
    王培林
    陈研
    王英杰
    魏龙
    单保慈
    王宝义
    中国物理C, 2011, (12) : 1162 - 1165