BROAD-BAND EXTREME-ULTRAVIOLET SURVEY SPECTROMETER FOR SHORT-TIME-SCALE EXPERIMENTS

被引:1
作者
CHAPMAN, BE [1 ]
DENHARTOG, DJ [1 ]
FONCK, RJ [1 ]
机构
[1] UNIV WISCONSIN, DEPT NUCL ENGN & ENGN PHYS, MADISON, WI 53706 USA
来源
APPLIED OPTICS | 1994年 / 33卷 / 19期
关键词
EUV SPECTROMETER; FLAT FOCAL FIELD; FIBEROPTIC COUPLER; PHOTODIODE ARRAY; SYNCHRONIZATION; CONTROL CIRCUIT;
D O I
10.1364/AO.33.004214
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A fast and inexpensive spectrometer system has been developed to record extreme-UV impurity spectra in a magnetic-fusion-research device. To simplify the vacuum system, light is passed out of the spectrometer's vacuum to the detector with a sodium-salicylate-coated, fiber-optic coupler. This coupler is positioned so that the focal field is nearly flat over its aperture. The system's detector is a microchannel-plate-intensified, linear, self-scanning photodiode array. The 1024-pixel array covers a bandwidth greater than 80 nm and is read out once every millisecond. The readout, which is four times faster than the manufacturer's maximum rating, is fully synchronized to the experiment by a locallY designed control circuit.
引用
收藏
页码:4214 / 4218
页数:5
相关论文
共 11 条
[1]   HIGH-RESOLUTION BENT-CRYSTAL SPECTROMETER FOR THE ULTRASOFT X-RAY REGION [J].
BEIERSDORFER, P ;
VONGOELER, S ;
BITTER, M ;
HILL, KW ;
HULSE, RA ;
WALLING, RS .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (05) :895-906
[2]   TIME-RESOLVING EXTREME ULTRAVIOLET SPECTROGRAPH FOR FUSION DIAGNOSTICS [J].
BELL, RE ;
FINKENTHAL, M ;
MOOS, HW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (12) :1806-1813
[3]   THE MADISON SYMMETRICAL TORUS [J].
DEXTER, RN ;
KERST, DW ;
LOVELL, TW ;
PRAGER, SC ;
SPROTT, JC .
FUSION TECHNOLOGY, 1991, 19 (01) :131-139
[4]   MULTICHANNEL GRAZING-INCIDENCE SPECTROMETER FOR PLASMA IMPURITY DIAGNOSIS - SPRED [J].
FONCK, RJ ;
RAMSEY, AT ;
YELLE, RV .
APPLIED OPTICS, 1982, 21 (12) :2115-2123
[5]  
FONCK RJ, 1983, ACS SYM SER, V236, P277
[6]   GRAZING-INCIDENCE TIME-RESOLVING SPECTROGRAPH FOR MAGNETIC FUSION PLASMA DIAGNOSTICS [J].
HODGE, WL ;
STRATTON, BC ;
MOOS, HW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1984, 55 (01) :16-24
[7]  
REGISTE DF, 1986, GAA18242 GEN AT TECH
[8]   HIGH-RESOLUTION POSITION-SENSITIVE DETECTOR FOR ULTRAVIOLET AND X-RAY PHOTONS [J].
RIEGLER, GR ;
MORE, KA .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1973, NS20 (01) :102-106
[9]   PHOTOELECTRON COUNTING WITH AN IMAGE INTENSIFIER TUBE AND A SELF-SCANNED PHOTODIODE ARRAY [J].
SANDEL, BR ;
BROADFOOT, AL .
APPLIED OPTICS, 1976, 15 (12) :3111-3114
[10]   NOISE IN LARGE-APERTURE SELF-SCANNED DIODE-ARRAYS [J].
SIMPSON, RW .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1979, 50 (06) :730-732