THE USE OF LASER DIFFRACTION FOR THE EVALUATION OF THE AEROSOL CLOUDS GENERATED BY MEDICAL NEBULIZERS

被引:124
作者
CLARK, AR
机构
[1] Genentech, Inc., South San Francisco
关键词
NEBULIZER; PARTICLE SIZE; DEPOSITION PROFILE;
D O I
10.1016/0378-5173(94)00255-4
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The diffraction method for the measurement of the droplet size distributions produced by medical nebulizers was investigated. The technique was found to be robust and reliable. Correlation of published results obtained from laser diffraction droplet size analysis to characterize the nebulizer clouds with radiolabel deposition profiles and predictions of an empirical deposition model, shows that the diffraction technique measures a size parameter relevant to the clinical situation. The high degree of correlation makes laser diffraction suitable for the predictive assessment of aerosols generated by nebulization of drug solutions.
引用
收藏
页码:69 / 78
页数:10
相关论文
共 17 条
[1]  
CIPOLLA DC, 1994, STP PHARMA SCI, V4, P50
[2]   EFFECT OF NEBULIZED AEROSOL SIZE ON LUNG DEPOSITION IN PATIENTS WITH MILD ASTHMA [J].
CLAY, MM ;
CLARKE, SW .
THORAX, 1987, 42 (03) :190-194
[3]  
Clifford R.H., 1990, CHEMISTRY, V62, P309
[4]  
GONDA I, 1990, AEROSOLS SCI IND HLT, V1, P227
[5]   LUNG DEPOSITION FROM 4 NEBULIZERS [J].
HARDY, JG ;
NEWMAN, SP ;
KNOCH, M .
RESPIRATORY MEDICINE, 1993, 87 (06) :461-465
[6]  
HO KKL, 1988, J PHARM PHARM S, V40, pP12
[7]  
HO KKL, 1986, J PHARM PHARMACOL, V38, pP26
[8]  
JOHNSON MA, 1989, CHEST, V96, P1
[9]  
KERKER M, 1989, SCATTERING LIGHT OTH, P356
[10]   PULMONARY DEPOSITION OF NEBULIZED PENTAMIDINE ISETHIONATE - EFFECT OF NEBULIZER TYPE, DOSE, AND VOLUME OF FILL [J].
ODOHERTY, M ;
THOMAS, S ;
PAGE, C ;
BRADBEER, C ;
NUNAN, T ;
BATEMAN, N .
THORAX, 1990, 45 (06) :460-464