The design and physical characterization of a multileaf collimator for robotic radiosurgery

被引:36
作者
Asmerom, G. [1 ]
Bourne, D. [1 ]
Chappelow, J. [1 ]
Goggin, L. M. [1 ]
Heitz, R. [1 ]
Jordan, P. [1 ]
Kilby, W. [1 ]
Laing, T. [1 ]
Maurer, C. R., Jr. [1 ]
Noll, J. M. [1 ]
Sayeh, S. [1 ]
Weber, A. [1 ]
机构
[1] Accuray Inc, 1310 Chesapeake Terrace, Sunnyvale, CA 94089 USA
关键词
x-ray beam collimation; CyberKnife (R) System; robotic radiosurgery; stereotactic body radiation therapy; multileaf collimator;
D O I
10.1088/2057-1976/2/1/017003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Amultileaf collimator (MLC) optimized for SBRT delivery with the CyberKnife (R) Robotic Radiosurgery System (Accuray Incorporated, Sunnyvale, CA, USA) is described. The MLC is exchangeable with the alternate fixed and variable circular aperture collimator systems. The non-coplanar workspace is effectively equivalent for all three collimation types. The same range of tracking options, including real-time respiratory motion tracking, and the same tolerance on beam pointing accuracy (<= 0.95mm) is maintained with all three collimation types. The MLC includes 52 flat-sided leaves, each of which is 90 mm tall and projects 3.85 mm width at the nominal treatment distance of 800 mm SAD. The design allows 100% overtravel and unrestricted interdigitation. Leaf position is determined by primary motor encoders and is checked with a secondary optical camera system. Maximum leakage, including inter-leaf and under the closed position leaf-tip gap was measured on five units to be <= 0.44%, while mean leakage and transmission ranged from 0.22%-0.25%. Leaf positioning accuracy measured over the full range of leaf positions, all robot and MLC orientations, and including variation with leaf motion direction and accumulated leaf motion after initialization had a mean error <0.2 mm, with 2%-98% range of <=+/- 0.5mm(projected at 800 mm SAD) on three units tested. The only factor found to effect leaf positioning accuracy was sag under gravity, which systematically altered leaf positions by 0.1 mm. Tilting the leaves to reduce inter-leaf leakage results in 0.5 mm asymmetry in leaf-side penumbra at 100 mm depth, and a partial leaf-edge transmission pattern analogous to the tongue and groove effect observed with interlocking leaves.
引用
收藏
页数:12
相关论文
共 13 条
[1]   Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system [J].
Antypas, Christos ;
Pantelis, Evaggelos .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (17) :4697-4718
[2]   MLC quality assurance techniques for IMRT applications [J].
Bayouth, JE ;
Wendt, D ;
Morrill, SM .
MEDICAL PHYSICS, 2003, 30 (05) :743-750
[3]   The characterization of unflattened photon beams from a 6 MV linear accelerator [J].
Cashmore, Jason .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (07) :1933-1946
[4]   An anthropomorphic phantom study of the accuracy of CyberKnife spinal radiosurgery [J].
Cheng, Y ;
Main, W ;
Taylor, D ;
Kuduvalli, G ;
Apuzzo, MLJ ;
Adler, JR ;
Wang, MY .
NEUROSURGERY, 2004, 55 (05) :1138-1146
[5]   The design, physical properties and clinical utility of an iris collimator for robotic radiosurgery [J].
Echner, G. G. ;
Kilby, W. ;
Lee, M. ;
Earnst, E. ;
Sayeh, S. ;
Schlaefer, A. ;
Rhein, B. ;
Dooley, J. R. ;
Lang, C. ;
Blanck, O. ;
Lessard, E. ;
Maurer, C. R., Jr. ;
Schlegel, W. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (18) :5359-5380
[6]   Characteristics and Performance Evaluation of the First Commercial MLC for a Robotic Delivery System [J].
Fuerweger, C. ;
Prins, P. ;
Coskan, H. ;
Heijmen, B. .
MEDICAL PHYSICS, 2015, 42 (06) :3194-3194
[7]   Dosimetric Comparison Between 3-Dimensional Conformal and Robotic SBRT Treatment Plans for Accelerated Partial Breast Radiotherapy [J].
Goggin, L. M. ;
Descovich, M. ;
McGuinness, C. ;
Shiao, S. ;
Pouliot, J. ;
Park, C. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2016, 15 (03) :437-445
[8]   Dosimetric characterization of a new miniature multileaf collimator [J].
Harmann, GH ;
Föhlisch, F .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (12) :N171-N177
[9]   Acceptance Testing and Commissioning Measurements of a Newly Released InCisea,, c Multileaf Collimator for CyberKnife M6a,, c System [J].
Huq, M. Saiful ;
Ozhasoglu, C. ;
Jang, S. ;
Hwang, M. ;
Heron, D. ;
Lalonde, R. .
MEDICAL PHYSICS, 2015, 42 (06) :3351-3351
[10]   The CyberKnife® Robotic Radiosurgery System in 2010 [J].
Kilby, W. ;
Dooley, J. R. ;
Kuduvalli, G. ;
Sayeh, S. ;
Maurer, C. R., Jr. .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2010, 9 (05) :433-452