AUTOMATION AND PROTHROMBIN TIME - A UNITED-KINGDOM FIELD-STUDY OF 2 WIDELY USED COAGULOMETERS

被引:48
作者
THOMSON, JM [1 ]
TABERNER, DA [1 ]
POLLER, L [1 ]
机构
[1] WITHINGTON HOSP,UNITED KINGDOM REFERENCE LAB ANTICOAGULANT REAGENTS & CONTROL,MANCHESTER M20 8LR,LANCS,ENGLAND
关键词
D O I
10.1136/jcp.43.8.679
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Current performance in the prothrombin time (PT) of the two main United Kingdom coagulometer/thromboplastin systems was assessed in a field survey. Twenty abnormal samples covering a wide spectrum of International Normalised Ratio (INR) were distributed to users of the KC4/KC1O and Coag-a-Mate instruments. Coagulometer results were compared with those of the manual method. A substantial minority with each system showed good agreement with the manual reference values. There was, however, a considerable variation between instruments, meaningful in clinical terms, evidenced by varying regression slopes and local system International Sensitivity Indices (ISI). For intense anticoagulation (3·0 to 4·5 INR) a larger dose of warfarin is needed with the Coag-a-Mate than with the KC instruments. With a manual INR of 4·0 the KC instruments tended to give longer PT (mean INR + 03); the Coag-a-Mate PT was generally shorter (mean INR-0·1). With both systems the mean normal PT were shorter than the manual but the degree of shortening did not parallel that of the abnormal samples. This effect undermines the use of a simple prothrombin ratio and of an INR value derived from it, based on a manual ISI. The use of a system related ISI cannot, however, be recommended until local instrument variables are controlled.
引用
收藏
页码:679 / 684
页数:6
相关论文
共 6 条
  • [1] COMPARISON OF 2 AUTOMATED COAGULOMETERS AND THE MANUAL TILT-TUBE METHOD FOR THE DETERMINATION OF PROTHROMBIN TIME
    DANGELO, A
    SEVESO, MP
    DANGELO, SV
    GILARDONI, F
    MACAGNI, A
    MANOTTI, C
    BONINI, P
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1989, 92 (03) : 321 - 328
  • [2] EFFECT OF AUTOMATION ON PROTHROMBIN TIME TEST IN NEQAS SURVEYS
    POLLER, L
    THOMSON, JM
    TABERNER, DA
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1989, 42 (01) : 97 - 100
  • [3] SURVEY OF PROTHROMBIN TIME IN NATIONAL-EXTERNAL-QUALITY-ASSESSMENT-SCHEME EXERCISES (1980-87)
    POLLER, L
    TABERNER, DA
    THOMSON, JM
    DARBY, KV
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1988, 41 (04) : 361 - 364
  • [4] EFFECT OF INTERNATIONAL SENSITIVITY INDEX (ISI) OF THROMBOPLASTINS ON PRECISION OF INTERNATIONAL NORMALIZED RATIOS (INR)
    TABERNER, DA
    POLLER, L
    THOMSON, JM
    DARBY, KV
    [J]. JOURNAL OF CLINICAL PATHOLOGY, 1989, 42 (01) : 92 - 96
  • [5] 1983, WHO TECHNICAL REPORT, P1
  • [6] 1984, WHO TECHNICAL REPORT, P19