ANALYSIS OF DEZOCINE IN SERUM AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND PRECOLUMN DERIVATIZATION

被引:1
作者
WILSON, JM
COHEN, RI
KEZER, EA
SMITH, ER
机构
[1] UNIV MASSACHUSETTS,MED CTR,DEPT ANESTHESIOL,WORCESTER,MA 01655
[2] UNIV MASSACHUSETTS,MED CTR,DEPT PHARMACOL,WORCESTER,MA 01655
来源
JOURNAL OF LIQUID CHROMATOGRAPHY | 1994年 / 17卷 / 19期
关键词
D O I
10.1080/10826079408013614
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We present a reverse phase liquid chromatographic assay involving pre-column o-phthalaldehyde derivatization and fluorescence detection with the capability of measuring dezocine in human serum and urine for pharmacokinetic or forensic purposes. It features a simple liquid-liquid extraction scheme using hexane, conservative detection limits of 1 ng/ml, linearity to 1000 ng/ml, precision of 10% (CV%) in the low ng/ml region, approximate chromatography time of 25 minutes and the use of a new internal standard, WY-19083.
引用
收藏
页码:4245 / 4257
页数:13
相关论文
共 50 条
  • [21] DETERMINATION OF PSEUDOURIDINE IN HUMAN URINE AND SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH POSTCOLUMN FLUORESCENCE DERIVATIZATION
    UMEGAE, Y
    NOHTA, H
    OHKURA, Y
    [J]. JOURNAL OF CHROMATOGRAPHY, 1990, 515 : 495 - 501
  • [22] PRECOLUMN COULOMETRIC CELL FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    SCHIEFFER, GW
    [J]. ANALYTICAL CHEMISTRY, 1981, 53 (01) : 126 - 127
  • [23] DETERMINATION OF PHENYLPROPANOLAMINE IN SERUM AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    DOWSE, R
    HAIGH, JM
    KANFER, I
    [J]. JOURNAL OF PHARMACEUTICAL SCIENCES, 1983, 72 (09) : 1018 - 1020
  • [24] DETERMINATION OF FUROSEMIDE IN SERUM AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    KUBO, H
    LI, H
    KOBAYASHI, Y
    KINOSHITA, T
    [J]. BUNSEKI KAGAKU, 1986, 35 (03) : 259 - 262
  • [25] DETERMINATION OF SPARFLOXACIN IN SERUM AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    BORNER, K
    BORNER, E
    LODE, H
    [J]. JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1992, 579 (02): : 285 - 289
  • [26] QUANTITATION OF IOTHALAMATE IN SERUM AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    WEBER, AF
    LEE, DW
    OPHEIM, K
    SMITH, AL
    [J]. JOURNAL OF CHROMATOGRAPHY, 1985, 337 (02): : 434 - 440
  • [27] FLUORIMETRIC DERIVATIZATION IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    LAWRENCE, JF
    [J]. JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1979, 17 (03) : 147 - 151
  • [28] DETERMINATION OF PSEUDOURIDINE AND 5-FLUOROPYRIMIDINES IN HUMAN-SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH PRECOLUMN FLUOROMETRIC DERIVATIZATION
    YOSHIDA, S
    ADACHI, T
    HIROSE, S
    [J]. MICROCHEMICAL JOURNAL, 1989, 39 (03) : 351 - 360
  • [29] QUANTIFYING ETHANOL BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH PRECOLUMN ENZYMATIC CONVERSION AND DERIVATIZATION WITH FLUOROMETRIC DETECTION
    CHEN, HM
    PETERSON, CM
    [J]. ALCOHOL, 1994, 11 (06) : 577 - 582
  • [30] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF GUANIDINO COMPOUNDS USING BENZOIN AS A PRECOLUMN FLUORESCENT DERIVATIZATION REAGENT
    KAI, M
    MIYAZAKI, T
    YAMAGUCHI, M
    OHKURA, Y
    [J]. JOURNAL OF CHROMATOGRAPHY, 1983, 268 (03): : 417 - 424