MAGNETIC RESONANCE IMAGING;
WATER DIFFUSION;
APPARENT RESTRICTION;
RAT BRAIN;
D O I:
10.1002/mrm.1910320520
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
The application of bipolar diffusion sensitizing gradient pulses to significantly reduce the diffusion time is described. This approach is combined with the rapid U-FLARE imaging sequence. Three diffusion-sensitized types of experiments are compared and their suitability for detecting restricted diffusion is discussed. Experiments using a modification of the diffusion weighting by varying the diffusion time between 1.6 and 6.0 ms obtained nonmonoexponential signal attenuation curves from both healthy brains and postmortem. This behavior is indicative of restricted diffusion, but as it is detectable only at short diffusion times, in contrast to a restriction due to impermeable barriers, we have termed this ''apparent restriction.''