VALIDATION OF A LIQUID-CHROMATOGRAPHY POSTCOLUMN DERIVATIZATION ASSAY FOR THE DETERMINATION OF CISPLATIN IN PLASMA

被引:15
作者
FARRISH, HH
HSYU, PH
PRITCHARD, JF
BROUWER, KR
JARRETT, J
机构
[1] Glaxo Inc. Research Institute, Research Triangle Park, NC 27709, Five Moore Drive
关键词
CISPLATIN; PLASMA; VALIDATED HPLC ASSAY; POSTCOLUMN DERIVATIZATION; UV DETECTION;
D O I
10.1016/0731-7085(94)90038-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Method validation results are described for a cisplatin LC post-column derivatization assay. Cisplatin plasma samples were treated with acetonitrile and a citrate buffer solution to enhance cisplatin stability. Processed samples were analysed on a chemically generated anion exchange column using a customized post-column derivatization platform and refrigerated autosampler. The UV response was monitored at 290 nm. The retention time of cisplatin was 9 min. The assay was linear from 0.06 to 30.0 mug ml-1 (r > = 0.998) with inter-run precisions (RSD) of 8.2% (n = 8), 5.9% (n = 8) and 4.0% (n = 8) for low (0.18 mug ml-1), medium (1.5 mug ml-1) and high (24.0 mug ml-1) quality control samples, respectively. The validated assay was used to monitor cisplatin levels in cisplatin drug interaction studies.
引用
收藏
页码:265 / 271
页数:7
相关论文
共 8 条
[1]   A HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY WITH IMPROVED SELECTIVITY FOR CISPLATIN AND ACTIVE PLATINUM(II) COMPLEXES IN PLASMA ULTRAFILTRATE [J].
ANDREWS, PA ;
WUNG, WE ;
HOWELL, SB .
ANALYTICAL BIOCHEMISTRY, 1984, 143 (01) :46-56
[2]   HOSPITAL ROUTINE ANALYSIS OF PENICILLINS, 3RD-GENERATION CEPHALOSPORINS AND AZTREONAM BY CONVENTIONAL AND HIGH-SPEED HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
JEHL, F ;
BIRCKEL, P ;
MONTEIL, H .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 413 :109-119
[3]   HPLC DETERMINATION OF CIS-DIAMMINEDICHLOROPLATINUM(II) IN PLASMA AND URINE WITH UV DETECTION AND COLUMN-SWITCHING [J].
KIZU, R ;
HAYAKAWA, K ;
MIYAZAKI, M .
BIOMEDICAL CHROMATOGRAPHY, 1989, 3 (01) :14-19
[4]   THE REACTIVITY OF CISPLATIN IN PLASMA - IMPLICATIONS FOR SAMPLE HANDLING IN PHARMACOKINETIC STUDIES [J].
LONG, DF ;
REPTA, AJ ;
STERNSON, LA .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1980, 6 (02) :167-173
[5]   POST-COLUMN REACTION DETECTOR FOR PLATINUM(II) ANTINEOPLASTIC AGENTS [J].
MARSH, KC ;
STERNSON, LA ;
REPTA, AJ .
ANALYTICAL CHEMISTRY, 1984, 56 (03) :491-497
[6]  
Repta A.J., 1980, CISPLATIN, P285, DOI [10.1016/b978-0-12-565050-2.50022-x, DOI 10.1016/B978-0-12-565050-2.50022-X]
[7]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY OF CIS-DICHLORODIAMMINEPLATINUM(II) USING CHEMICALLY-BONDED AND SOLVENT-GENERATED ION-EXCHANGERS [J].
RILEY, CM ;
STERNSON, LA ;
REPTA, AJ .
JOURNAL OF CHROMATOGRAPHY, 1981, 217 (NOV) :405-420
[8]  
STERNSON LA, 1983, ANAL P, V20, P366