PET performance evaluation of the small-animal Hyperion IID PET/MRI insert based on the NEMA NU-4 standard

被引:30
作者
Hallen, Patrick [1 ]
Schug, David [1 ]
Weissler, Bjoern [1 ]
Gebhardt, Pierre [1 ]
Salomon, Andre [3 ]
Kiessling, Fabian [2 ]
Schulz, Volkmar [1 ,3 ]
机构
[1] Rhein Westfal TH Aachen, Inst Expt Mol Imaging, Dept Phys Mol Imaging Syst, Aachen, Germany
[2] Rhein Westfal TH Aachen, Inst Expt Mol Imaging, Aachen, Germany
[3] Philips Res, Dept Oncol Solut, Eindhoven, Netherlands
基金
英国惠康基金;
关键词
positron emission tomography; small-animal imaging; performance evaluation; simultaneous PET/MRI;
D O I
10.1088/2057-1976/aae6c2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Hyperion IID PET insert is the first scanner using fully digital silicon photomultipliers for simultaneous PET/MR imaging of small animals up to rabbit size. In this work, we evaluate the PET performance based on the National Eletrical Manufacturers Association (NEMA) NU 4-2008 standard, whose standardized measurement protocols allow comparison of different small-animal PET scanners. The Hyperion IID small-animal PET/MR insert comprises three rings of 20 detector stacks with pixelated scintillator arrays with a crystal pitch of 1 mm, read out with digital silicon photomultipliers. The scanner has a large ring diameter of 209.6 mm and an axial field of view of 96.7 mm. We evaluated the spatial resolution, energy resolution, time resolution and sensitivity by scanning a 22Na point source. The count rates and scatter fractions were measured for a wide range of F-18 activity inside a mouse-sized scatter phantom. We evaluated the image quality using the mouse sized image quality phantom specified in the NEMA NU4 standard, filled with F-18. Additionally, we verified the in-vivo imaging capabilities by performing a simultaneous PET/MRI scan of a mouse injected with F-18-FDG. We processed all measurement data with an energy window of 250 keV to 625 keV and a coincidence time window of 2 ns. The filtered-backprojection reconstruction of the point source has a full width at half maximum (FWHM) of 1.7 mm near the isocenter and degrades to 2.5 mm at a radial distance of 50 mm. The scanner's average energy resolution is 12.7% (AE/E FWHM) and the coincidence resolution time is 609 ps. The peak absolute sensitivity is 4.0% and the true and noise-equivalent count rates reach their peak at an activity of 46 MBq with 483 kcps and 407 kcps, respectively, with a scatter fraction of 13%. The iterative reconstruction of the image quality phantom has a uniformity of 3.7%, and recovery coefficients from 0.29, 0.91 and 0.94 for rod diameters of 1 mm, 3 mm and 5 mm, respectively. After application of scatter and attenuation corrections, the air- and water-filled cold regions have spill-over ratios of 6.3% and 5.4%, respectively. The Hyperion IID PET/MR insert provides state-of-the-art PET performance while enabling simultaneous PET/MRI acquisition of small animals up to rabbit size.
引用
收藏
页数:12
相关论文
共 40 条
[1]   Study design in MRI: Basic principles [J].
Amaro, E ;
Barker, GJ .
BRAIN AND COGNITION, 2006, 60 (03) :220-232
[2]  
ANDREYEV A, 2017, J NUCL MED S1, V58
[3]   Performance Evaluation of the Inveon Dedicated PET Preclinical Tomograph Based on the NEMA NU-4 Standards [J].
Bao, Qinan ;
Newport, Danny ;
Chen, Mu ;
Stout, David B. ;
Chatziioannou, Arion F. .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (03) :401-408
[4]  
Berger MJ, 2010, NON TRADITIONAL REF
[5]   Performance Evaluation of the LabPET12, a Large Axial FOV APD-based Digital PET Scanner [J].
Bergeron, M. ;
Cadorette, J. ;
Bureau-Oxton, C. ;
Beaudoin, J. -F. ;
Tetrault, M. -A. ;
Leroux, J. -D. ;
Lepage, M. D. ;
Robert, G. ;
Fontaine, R. ;
Lecomte, R. .
2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, :4017-+
[6]   High-Resolution Underwater Mapping Using Side-Scan Sonar [J].
Burguera, Antoni ;
Oliver, Gabriel .
PLOS ONE, 2016, 11 (01)
[7]   NEMA NU 4-2008 Performance Measurements of Two Commercial Small-Animal PET Scanners: ClearPET and rPET-1 [J].
Canadas, Mario ;
Embid, Miguel ;
Lage, Eduardo ;
Desco, Manuel ;
Jose Vaquero, Juan ;
Manuel Perez, Jose .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (01) :58-65
[8]   Motion Correction Options in PET/MRI [J].
Catana, Ciprian .
SEMINARS IN NUCLEAR MEDICINE, 2015, 45 (03) :212-223
[9]   MRI-Based Nonrigid Motion Correction in Simultaneous PET/MRI [J].
Chun, Se Young ;
Reese, Timothy G. ;
Ouyang, Jinsong ;
Guerin, Bastien ;
Catana, Ciprian ;
Zhu, Xuping ;
Alpert, Nathaniel M. ;
El Fakhri, Georges .
JOURNAL OF NUCLEAR MEDICINE, 2012, 53 (08) :1284-1291
[10]  
DAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717