INFLUENCE OF ELUTION CONDITIONS ON HPLC RETENTION INDEX VALUES OF SELECTED ACIDIC AND BASIC DRUGS MEASURED IN THE 1-NITROALKANE SCALE

被引:14
作者
BOGUSZ, M [1 ]
机构
[1] UNIV HEIDELBERG,INST FORENS MED,W-6900 HEIDELBERG,GERMANY
关键词
D O I
10.1093/jat/15.4.174
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The retention indices (RI) of 62 acidic, neutral, and basic drugs were determined by HPLC under isocratic conditions (20, 40, and 60% acetonitrile-phosphate buffer, pH 3.2, containing 0.05% nonylamine) and with gradient elution (5-70% acetonitrile-phosphate buffer pH 3.2, containing 0.05% nonylamine). The RI values of acidic drugs and earlier-eluting basic drugs decreased distinctly with increasing concentrations of acetonitrile. For the later-eluting basic drugs, the opposite trend was observed. The RI data obtained with different isocratic conditions and with gradient elution were not transferable. The study shows the importance of careful standardization of elution conditions for interlaboratory comparison.
引用
收藏
页码:174 / 178
页数:5
相关论文
共 25 条
[1]   NITROALKANES AS A MULTIDETECTOR RETENTION INDEX SCALE FOR DRUG IDENTIFICATION IN GAS-CHROMATOGRAPHY [J].
ADERJAN, R ;
BOGUSZ, M .
JOURNAL OF CHROMATOGRAPHY, 1988, 454 :345-351
[2]   GAS-LIQUID-CHROMATOGRAPHIC RETENTION INDEXES OF 1318 SUBSTANCES OF TOXICOLOGICAL INTEREST ON SE-30 OR OV-1 STATIONARY PHASE [J].
ARDREY, RE ;
MOFFAT, AC .
JOURNAL OF CHROMATOGRAPHY, 1981, 220 (03) :195-252
[3]   RETENTION INDEX SCALE FOR LIQUID-LIQUID CHROMATOGRAPHY [J].
BAKER, JK ;
MA, CY .
JOURNAL OF CHROMATOGRAPHY, 1979, 169 (FEB) :107-115
[4]   LABORATORY TO LABORATORY REPRODUCIBILITY OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC RETENTION INDEXES [J].
BAKER, JK ;
CATES, LA ;
CORBETT, MD ;
HUBER, JW ;
LATTIN, DL .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1982, 5 (05) :829-839
[5]   SURFACE SILANOLS IN SILICA-BONDED HYDROCARBONACEOUS STATIONARY PHASES .2. IRREGULAR RETENTION BEHAVIOR AND EFFECT OF SILANOL MASKING [J].
BIJ, KE ;
HORVATH, C ;
MELANDER, WR ;
NAHUM, A .
JOURNAL OF CHROMATOGRAPHY, 1981, 203 (JAN) :65-84
[6]   CORRECTION OF RETENTION INDEX VALUES IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AS A TOOL FOR COMPARISON OF RESULTS OBTAINED WITH DIFFERENT OCTADECYL SILICA PHASES [J].
BOGUSZ, M .
JOURNAL OF CHROMATOGRAPHY, 1987, 387 :404-409
[7]  
BOGUSZ M, 1987, Z RECHTSMED, V98, P263
[8]   USE OF CORRECTED RETENTION INDEXES BASED ON 1-NITROALKANE AND ALKYL ARYLKETONE SCALES FOR HPLC IDENTIFICATION OF BASIC DRUGS [J].
BOGUSZ, M ;
NEIDLFISCHER, G ;
ADERJAN, R .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1988, 12 (06) :325-329
[9]  
BOGUSZ M, 1991, IN PRESS 26TH P TIAF
[10]  
BOGUSZ M, 1988, J ANAL TOXICOL, V12, P62