A SIMPLE, RAPID METHOD FOR THE SIMULTANEOUS DETERMINATION OF MORPHINE AND ITS PRINCIPAL METABOLITES IN PLASMA USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND FLUOROMETRIC DETECTION

被引:67
|
作者
GLARE, PA
WALSH, TD
PIPPENGER, CE
机构
[1] CLEVELAND CLIN FDN,DEPT HEMATOL & MED ONCOL,PALLIAT CARE SERV,DESK T-33,1 CLIN CTR,CLEVELAND,OH 44195
[2] CLEVELAND CLIN FDN,DEPT BIOCHEM,APPL CLIN PHARMACOL SECT,CLEVELAND,OH 44195
关键词
MORPHINE; METABOLITES; HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY; FLUOROMETRY; PLASMA;
D O I
10.1097/00007691-199105000-00007
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
This article describes a high-performance liquid chromatography (HPLC) method for the simultaneous determination of morphine (M) and its principal metabolites morphine-3-glucuronide (M3G), morphine-6-glucuronide (M6G), and normorphine (NM) in plasma. All four compounds are extracted from plasma using a C8 solid-phase extraction column, separated by reverse-phase HPLC on a C18 analytical column, and detected by spectrofluorometry at 210 nm excitation wavelength. The method takes advantage of the compounds' native fluorescence, so that derivitization is not required. Samples have been quantified over a concentration range of 25-100 ng/ml M and NM, 50-200 ng/ml M3G, and 100-300 ng/ml M6G, using nalorphine (500 ng/ml) as internal standard. Within-run and between-run errors were < 10% for morphine and < 13% for all the metabolites. The lower limit of quantitation for morphine is 10 ng/ml. The accuracy of the method was confirmed by including quality controls fitted to the standard curves of each compound. The assay described in this article represents a simplification of previous versions of the method, which included cumbersome extraction procedures and multiple detectors. For the first time, an internal standard has been employed. The assay is reliable and easy to use and can be performed in any therapeutic drug monitoring laboratory.
引用
收藏
页码:226 / 232
页数:7
相关论文
共 50 条
  • [41] Simultaneous determination of tricarboxylic acid cycle metabolites by high-performance liquid chromatography with ultraviolet detection
    Shurubor, Yevgeniya I.
    Cooper, Arthur J. L.
    Isakova, Elena P.
    Deryabina, Yulia I.
    Beal, M. Flint
    Krasnikov, Boris F.
    ANALYTICAL BIOCHEMISTRY, 2016, 503 : 8 - 10
  • [42] Quantitative determination of tolbutamide and its metabolites in human plasma and urine by high-performance liquid chromatography and UV detection
    Hansen, LL
    Brosen, K
    THERAPEUTIC DRUG MONITORING, 1999, 21 (06) : 664 - 671
  • [43] Rapid, simple, and clinically applicable high-performance liquid chromatography method for clinical determination of plasma colistin concentrations
    Yuki Hanai
    Kazuhiro Matsuo
    Takayoshi Kosugi
    Ayumu Kusano
    Hayato Ohashi
    Itsuki Kimura
    Shinobu Hirayama
    Yuta Nanjo
    Yoshikazu Ishii
    Takahiro Sato
    Taito Miyazaki
    Kenji Nishizawa
    Takashi Yoshio
    Journal of Pharmaceutical Health Care and Sciences, 4 (1)
  • [44] Rapid, simple, and clinically applicable high-performance liquid chromatography method for clinical determination of plasma colistin concentrations
    Hanai, Yuki
    Matsuo, Kazuhiro
    Kosugi, Takayoshi
    Kusano, Ayumu
    Ohashi, Hayato
    Kimura, Itsuki
    Hirayama, Shinobu
    Nanjo, Yuta
    Ishii, Yoshikazu
    Sato, Takahiro
    Miyazaki, Taito
    Nishizawa, Kenji
    Yoshio, Takashi
    JOURNAL OF PHARMACEUTICAL HEALTH CARE AND SCIENCES, 2018, 4
  • [45] DETERMINATION OF BACLOFEN IN HUMAN PLASMA AND URINE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH FLUORESCENCE DETECTION
    TOSUNOGLU, S
    ERSOY, L
    ANALYST, 1995, 120 (02) : 373 - 375
  • [46] SENSITIVE DETERMINATION OF CINNARIZINE IN HUMAN PLASMA BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND FLUORESCENCE DETECTION
    ROSSEEL, MT
    LEFEBVRE, RA
    CHROMATOGRAPHIA, 1993, 36 : 356 - 358
  • [47] Determination of morphine in the plasma of addicts in using Zeolite Y extraction following high-performance liquid chromatography
    Ghazi-Khansari, M
    Zendehdel, R
    Pirali-Hamedani, M
    Amini, M
    CLINICA CHIMICA ACTA, 2006, 364 (1-2) : 235 - 238
  • [48] A simple high-performance liquid chromatography assay for simultaneous measurement of adenosine, guanosine, and the oxypurine metabolites in plasma
    Feng, JDZ
    Yeung, PKF
    THERAPEUTIC DRUG MONITORING, 2000, 22 (02) : 177 - 183
  • [49] SIMPLE DETERMINATION OF HALOFUGINONE IN CHICKEN TISSUE BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY
    YAMAMOTO, Y
    HASHIGUCHI, R
    ARAKI, K
    KUSHIMA, H
    JOURNAL OF THE FOOD HYGIENIC SOCIETY OF JAPAN, 1991, 32 (05): : 444 - 447
  • [50] Simple determination of pirfenidone in rat plasma via high-performance liquid chromatography
    Wang, Yongsheng
    Zhao, Xiaoping
    Zhong, Jiao
    Chen, Yan
    Liu, Xiaoquan
    Wang, Guangji
    BIOMEDICAL CHROMATOGRAPHY, 2006, 20 (12) : 1375 - 1379