摘要
A system for low temperature nuclear orientation studies is presented where the sample can be prepared by in situ implantation of ions from an isotope separator into a host lattice at temperatures <0.5 K. The special construction of the cryostat necessary for this purpose as well as the beam line from the isotope separator to the cryostat are described in detail. A description of the data acquisition arrangement appropriate to the experiments planned is given. The cooling characteristics of the refrigerator under different operating conditions are discussed. A first successful nuclear magnetic resonance experiment on oriented 197mHg implanted in situ at T<0.5 K is reported. © 1978.
引用
收藏
页码:421 / 428
页数:8
相关论文
共 5 条
[1]   INSITU NMR-ON ON HG-197M IMPLANTED AT T LESS-THAN 0.5 K IN IRON [J].
HERZOG, P ;
FOLLE, HR ;
FREITAG, K ;
REUSCHENBACH, M ;
SCHMIDT, HU .
PHYSICS LETTERS A, 1977, 62 (04) :245-246
[2]   COMBINED LATTICE-LOCATION-HYPERFINE-INTERACTION EXPERIMENTS ON HG IMPLANTED IN FE [J].
JAMES, PK ;
HERZOG, P ;
STONE, NJ ;
FREITAG, K .
PHYSICAL REVIEW B, 1976, 13 (01) :59-67
[3]   MEASUREMENTS OF MAGNETIC HYPERFINE FIELDS AT HG IN FE, CO, AND NI HOSTS BY E--GAMMA TIME-DIFFERENTIAL PERTURBED-ANGULAR-CORRELATION METHOD [J].
KRIEN, K ;
BIBILONI, AG ;
FREITAG, K ;
SOARES, JC ;
VIANDEN, R .
PHYSICAL REVIEW B, 1973, 8 (05) :2248-2254
[4]  
Lounasmaa O. V., 1974, EXPT PRINCIPLES METH
[5]   PRECISION MAGNETIC-MOMENT DETERMINATIONS FOR 43-MIN HG-199M AND OTHER ISOMERS OF MERCURY [J].
REIMANN, RJ ;
MCDERMOTT, MN .
PHYSICAL REVIEW C, 1973, 7 (05) :2065-2079