Development of printed and flexible dry ECG electrodes

被引:138
|
作者
Chlaihawi, Amer Abdulmahdi [1 ,2 ]
Narakathu, Binu Baby [1 ]
Emamian, Sepehr [1 ]
Bazuin, Bradley J. [1 ]
Atashbar, Massood Z. [1 ]
机构
[1] Western Michigan Univ, Dept Elect & Comp Engn, 4601 Campus Dr, Kalamazoo, MI 49008 USA
[2] Univ Kufa, Dept Elect Engn, Najaf, Iraq
关键词
Dry ECG electrode; ECG monitoring; MWCNT; PDMS; Screen printing; Wearable biomedical sensors;
D O I
10.1016/j.sbsr.2018.05.001
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Printed, flexible and wearable dry electrodes for monitoring electrocardiogram (ECG) signals, without any skin preparation and use of wet gel, has been developed. Silver (Ag) flake ink was screen printed on a flexible polyethylene terephthalate (PET) substrate to fabricate the dry ECG electrode. Multi-walled carbon nanotube (MWCNT)/polydimethylsiloxane (PDMS) composite, as a conductive polymer, was then deposited on the printed Ag electrode by using a bar coating technique. The performance of the printed electrodes was investigated by testing the MWCNT/PDMS composite conductivity and measuring the electrode-skin impedance for electrode radii varying from 8mm to 16 mm. It was observed that the dry ECG electrode, with the largest area, demonstrated better performance, in terms of MWCNT/PDMS composite conductivity, ECG signal intensity and correlation when compared to a commercial wet silver/silver chloride (Ag/AgCl) electrode. In addition, the capability of the dry ECG electrodes for monitoring ECG signals in both the relaxed sitting position and while the subject is in motion, was also investigated and the results were compared with a wet Ag/AgCl ECG electrode (T716). While the subject is in motion, the printed dry electrodes were less noisy and were able to better identify the typical ECG characteristics in the signals due to its better conformal contact at the electrode-skin interface. The results obtained demonstrated the feasibility of employing conventional screen printing process for the development of flexible dry ECG electrodes for applications in the biomedical industry.
引用
收藏
页码:9 / 15
页数:7
相关论文
共 50 条
  • [31] Heart Monitor Using Flexible Capacitive ECG Electrodes
    Gao, Yang
    Soman, Varun V.
    Lombardi, Jack P.
    Rajbhandari, Pravakar Prasad
    Dhakal, Tara P.
    Wilson, Dale G.
    Poliks, Mark D.
    Ghose, Kanad
    Turner, James N.
    Jin, Zhanpeng
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2020, 69 (07) : 4314 - 4323
  • [32] MWCNT-COOH/PDMS based dry ECG electrodes for ambulatory ECG recordings
    Gandhi, Bani
    Raghava, N. S.
    MATERIALS LETTERS, 2022, 307
  • [33] 3D Printed Dry EEG Electrodes
    Krachunov, Sammy
    Casson, Alexander J.
    SENSORS, 2016, 16 (10)
  • [34] Short-Term Evaluation of Dry Electrodes Fabricated using Flexible Printed Circuits Processes for Bioimpedance Measurements
    Critcher, Shelby
    Freeborn, Todd J.
    IEEE SOUTHEASTCON 2020, 2020,
  • [35] Development of PDMS-based flexible dry type SEMG electrodes by micromachining technologies
    Jung, Jung Mo
    Cha, Doo Yeol
    Kim, Deok Su
    Yang, Hee Jun
    Choi, Kyo Sang
    Choi, Jong Myoung
    Chang, Sung Pil
    APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 2014, 116 (03): : 1395 - 1401
  • [36] Development of PDMS-based flexible dry type SEMG electrodes by micromachining technologies
    Jung Mo Jung
    Doo Yeol Cha
    Deok Su Kim
    Hee Jun Yang
    Kyo Sang Choi
    Jong Myoung Choi
    Sung Pil Chang
    Applied Physics A, 2014, 116 : 1395 - 1401
  • [37] Development Of A Flexible Printed Battery
    Crosby, JustOne
    Maddipatla, Dinesh
    Wu, Qingliu
    Bazuin, Bradley
    Stoops, Matthew
    Atashbar, Massood
    2021 IEEE INTERNATIONAL CONFERENCE ON ELECTRO INFORMATION TECHNOLOGY (EIT), 2021, : 401 - 404
  • [38] Electrical Performance of Inkjet Printed Flexible Cable for ECG Monitoring
    Wan, Qiansu
    Yang, Geng
    Chen, Qiang
    Zheng, Li-Rong
    2011 IEEE 20TH CONFERENCE ON ELECTRICAL PERFORMANCE OF ELECTRONIC PACKAGING AND SYSTEMS (EPEPS), 2011, : 231 - 234
  • [39] Printed, flexible electrochromic displays using interdigitated electrodes
    Coleman, JP
    Lynch, AT
    Madhukar, P
    Wagenknecht, JH
    SOLAR ENERGY MATERIALS AND SOLAR CELLS, 1999, 56 (3-4) : 395 - 418
  • [40] Flexible Fibrous Piezoelectric Sensors on Printed Silver Electrodes
    Son, Ho Yeon
    Park, Jae Sung
    Huang, Jonathan
    Kim, Jiseok
    Nam, Yoon Sung
    Kim, Woo Soo
    IEEE TRANSACTIONS ON NANOTECHNOLOGY, 2014, 13 (04) : 709 - 713