ANALYSIS OF SOLUTION RESIDUES BY GLOW-DISCHARGE MASS-SPECTROMETRY

被引:27
|
作者
BARSHICK, CM
DUCKWORTH, DC
SMITH, DH
机构
[1] Analytical Chemistry Division, Oak Ridge National Laboratory, Oak Ridge, 37831-6375, TN
关键词
D O I
10.1016/1044-0305(93)85041-U
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A technique for the analysis of microliter volumes of solution by glow discharge mass spectrometry (GDMS) has been successfully demonstrated. Cathode preparation involves mixing an aliquot of the sample solution with a pure conducting powder, followed by drying and pressing before conventional GDMS analysis. The analyte signal at the 100-ppm level was observed to be stable to better than 5% for the duration of the analysis (30-45 min). Internal and external reproducibilities were better than 5%, and the ion signal intensity was linear with concentration over at least four orders of magnitude. Quantification was demonstrated by means of user-defined relative sensitivity factors. Relative standard deviations were better than 15% for the elements investigated, with no reconcentration of the analyte.
引用
收藏
页码:47 / 53
页数:7
相关论文
共 50 条