A LEAD(II) ION-SELECTIVE ELECTRODE VIA A METAL-COMPLEX OF POLY(HYDROXAMIC ACID)

被引:14
|
作者
ANUAR, K
HAMDAN, S
机构
[1] Department of Chemistry, Faculty of Science and Environmental Studies, Universiti Pertanian Malaysia, 43400 UPM Serdang, Selangor
关键词
D O I
10.1016/0039-9140(92)80199-N
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new lead(II) electrode has been constructed with poly(hydroxamic acid) (PHXA) as the active material and silicone rubber as the supporting material. The electrode gave near Nernstian response over the concentration range 4 x 10(-5)-1 x 10(-2)M lead(II). The detection limit of the electrode is approximately 4 x 10(-6)M and the electrode works well in the pH range 4.5-6.0. The response time was 50-120 sec over the whole concentration range and the electrode has a working life of at least 4 weeks. Iron(III) severely poisoned the electrode membrane. Nickel(II) and mercury(II) gave very strong interference compared to copper(II), silver(I), cobalt(II), sodium(I), potassium(I), zinc(II) and cadmium(II) which gave some or little interference. Values determined with atomic absorption (AAS) and a commercial lead(II) electrode were in good agreement with those measured with the lead(II) electrode reported here.
引用
收藏
页码:1653 / 1656
页数:4
相关论文
共 50 条