FACTORS AFFECTING EMISSION COLLECTION BY SURGICAL SMOKE EVACUATORS

被引:17
作者
SMITH, JP
TOPMILLER, JL
SHULMAN, S
机构
[1] National Institute for Occupational Safety and Health, Cincinnati, Ohio
关键词
air flow; gas evacuation; nozzle flow;
D O I
10.1002/lsm.1900100303
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Gaseous and particulate emissions from laser surgery are often controlled by the use of smoke evacuators. Thus it is important that factors affecting the performance of these evacuators be understood. In this study, a tracer gas technique was used to examine a number of factors affecting the performance of smoke evacuators, including evacuator flow rate, distance from the evacuator nozzle to the surgical site, and direction and speed of external air flow in relation to nozzle flow. The tracer gas technique allowed the release of emissions to be visualized with infrared imaging and also allowed the collection efficiency to be assessed quantitatively. The results were demonstrated with the use of a surgical laser system. It was found that the collection efficiency of the smoke evacuator was affected by all the factors studied. Increasing the evacuator flow rate allowed the collection of emissions under conditions in which lower evacuator flow rates had less efficient collection. The speed and direction of external air flow affected collection of emissions greatly. If the air flow was in the same direction as the nozzle flow, efficient collection of emissions at a distance from the emission release point was observed. However, at other angles relative to nozzle flow, the efficiency degraded rapidly with distance from the emission release site. With the use of the surgical laser system, at laser powers of 30 W, the tracer system predicted the collection of emissions. At 60 and 100 W of laser power, higher external air flows and greater attention to nozzle positioning were necessary. Based on the results of this study, conclusions are drawn on how to improve the collection efficiency of smoke evacuators used in laser surgery. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:224 / 233
页数:10
相关论文
共 10 条
[1]   THE EFFECTS OF LASER SMOKE ON THE LUNGS OF RATS [J].
BAGGISH, MS ;
ELBAKRY, M .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1987, 156 (05) :1260-1265
[2]   PROTECTION OF THE RAT LUNG FROM THE HARMFUL EFFECTS OF LASER SMOKE [J].
BAGGISH, MS ;
BALTOYANNIS, P ;
SZE, E .
LASERS IN SURGERY AND MEDICINE, 1988, 8 (03) :248-253
[3]  
FEITAG L, 1988, LASER SURG MED, V7, P283
[4]   PAPILLOMAVIRUS IN THE VAPOR OF CARBON-DIOXIDE LASER-TREATED VERRUCAE [J].
GARDEN, JM ;
OBANION, MK ;
SHELNITZ, LS ;
PINSKI, KS ;
BAKUS, AD ;
REICHMANN, ME ;
SUNDBERG, JP .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1988, 259 (08) :1199-1202
[5]  
KOKOSA J, 1988, SPIE, P908
[6]  
MIHASHI W, 1981, 4TH P C INT SOC LAS
[7]   SMOKE FROM LASER-SURGERY - IS THERE A HEALTH-HAZARD [J].
NEZHAT, C ;
WINER, WK ;
NEZHAT, F ;
NEZHAT, C ;
FORREST, D ;
REEVES, WG .
LASERS IN SURGERY AND MEDICINE, 1987, 7 (04) :376-382
[8]   EVALUATION OF A SMOKE EVACUATOR USED FOR LASER-SURGERY [J].
SMITH, JP ;
MOSS, CE ;
BRYANT, CJ ;
FLEEGER, AK .
LASERS IN SURGERY AND MEDICINE, 1989, 9 (03) :276-281
[9]  
1988, CLIN LASER MONTHLY, V6, P61
[10]  
1908, ANSIZ1363 AM NAT STA